Root NIBPT
Calibration for NIBP
The Calibration option on the NIBP menu allows a qualified service professional to access
calibration settings and tools for the NIBP module. For more information, see NIBP Module
Calibration Test on page 150.
Note: This section is provided as a reference and intended for qualified service professionals
only.
Sounds
Use the Sounds screen to control the volume level of sounds and duration of audio pause for
Root.
Option
Description
Alarm
Sets the alarm volume level. Highest
Volume
Pulse Tone
Sets the pulse tone volume
Volume
level.
Sets the length of time that
the audible alarm remains
Audio Pause
silenced, when Audio Pause
Duration
is enabled. See Audio Pause
on page 80.
Allows the audible pulse to
continue to beep when the
SmartTone
pleth graph shows signs of
motion.
*Requires the all mute enabled option to be toggled to ON in the Access Control menu. See
Access Control on page 70.
www.masimo.com
Factory
Configurable Settings
Default
Setting
Slide towards the left to decrease
volume
volume to silence.
Slide towards the left to decrease
3
volume to silence.
1, 2, 3 minutes, Permanent*,
Permanent with Reminder*.
If Permanent is selected, there will
be no audible alarms when Audio
Pause is enabled, but visual
2 minutes
alarms will still display.
If Permanent with Reminder is
selected, a tone will sound every
three (3) minutes as a reminder
that Permanent is active when
Audio Pause is enabled.
Off
On or Off
63