Root NIBPT
Guidance and Manufacturer's Declaration-Electromagnetic
Immunity
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME Equipment should assure that it is used in such an
environment.
Immunity Test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines.
IEC 61000-4-11
Power frequency (50 /
60 Hz) magnetic field.
IEC 61000-4-8
www.masimo.com
IEC 60601
Compliance
Test Level
Level
+6 kV contact
+6 kV
contact
+8 kV air
+8 kV air
+2 kV for
power supply
lines.
+1 kV for
input/output
lines.
+1 kV -
differential
mode
+2 kV -
common mode
100% for 0.5
cycle
60% for 5
cycles
30% for 25
cycles
100% for 5
seconds
3 A/m
3 A/m
Electromagnetic Environment -
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
---
Mains power quality should be that
of a typical hospital environment.
---
Mains power quality should be that
of a typical commercial or hospital
environment. Root provides a
battery for continued operation
during power mains interruption for
---
a maximum of 4 hours.
Power frequency magnetic fields
should be at levels characteristic of
typical location in a typical hospital
environment.
142