Summary of Contents for Masimo Radical Signal Extraction

  • Page 1 signal extraction pulse oximeter O P E R A T O R ’ S M A N U A L...
  • Page 2 S i g n a l E x t r a c t i o n P u l s e O x i m e t e r O P E R A T O R ’ S M A N U A L...
  • Page 3 6,430,525, 6,360,114, 6,263,222, 6,236,872, 6,229,856, 6,206,830, 6,157,850, 6,067,462, 6,011,986, 6,002,952, 5,919,134, 5,823,950, 5,769,785, 5,758,644, 5,685,299, 5,632,272, 5,490,505, 5,482,036, international equivalents, or one or more of the patents referenced at www.masimo.com/patents. Products containing Satshare® feature are also covered by U.S. Patent 6,770,028. Other patents pending.
  • Page 4 ■ Do not place containers containing liquids on or near the pulse oximeter. Liquids spilled on the The Radical Signal Extraction Pulse Oximeter is designed to minimize the possibility of pulse oximeter may cause it to perform inaccurately or fail.
  • Page 5: Table Of Contents

    Menu tree ............4-11 Masimo SET signal extraction technology .......1-5 Alarms .
  • Page 6: About This Manual

    SYSTEM DESCRIPTION describes the Radical system and its Cleaning And Reuse Of Masimo Sensors .......8-4 functions and features.
  • Page 7: Product Description

    A WARNING is provided when actions may result in a serious outcome (i.e., injury, with a multiparameter patient monitor to provide Masimo SET pulse oximetry serious adverse affect, death) to the patient or user. Look for text in a gray shaded box.
  • Page 8: Pulse Oximetry

    LED's Recessed Photo Detector Once the Radical receives the signal from the sensor, it utilizes Masimo SET signal extraction technology for calculation of the patient’ s functional oxygen saturation and pulse rate. Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 9: Introduction

    ONLY USE A SATSHARE CABLE THAT HAS A FERRITE BEAD INSTALLED. Refer to Section 3, SatShare Setup and Section 4, SatShare Operation for additional details. *Contact Masimo for the latest list of SatShare validated multiparameter monitors. Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 10: Handheld Front Panel

    The following figure and corresponding text outline all the features of the Handheld Radical: calculated and the display is updated at a frequency of once per second. Pleth + Signal IQ View The Masimo SET symbol is shown on the Radical display when SET processing is active. Masimo MASIMO SET...
  • Page 11: Handheld Back Panel

    The pulse rate is calculated and the display is updated at a frequency of once per second. Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 12: Radical Standalone

    Docking Station Battery Deep ■ ■ ■ ■ ■ Discharge Support The RDS-1 and RDS-3 are optionally available with RadNet capability. (Refer to Section 3 RadNet Setup for details). Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 13: Standalone Back Panel

    CONNECTOR order of a physician (USA audiences only) that the correct cable is used for each type of patient monitor. Refer to the Masimo web site at www.masimo.com for the latest SatShare cables and validated instruments. Underwriter's Laboratories Inc. approved...
  • Page 14: Radical Monitor Interface

    ■ Upgrades any approved and validated monitor to Masimo SET performance by using The following sections of the manual describe the preparation, set-up and initial the calculated SpO and pulse rate determined by Radical to simulate an ideal installation of the Radical Pulse Oximeter.
  • Page 15: Initial Battery Charging

    Set to pre-power down setting LANGUAGE Set to pre-power down setting ANALOG OUTPUT Set to pre-power down setting SERIAL OUTPUT PORT MODE Set to pre-power down setting Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 16: Adult/Neo (Hospital Defined) Settings

    Select the SatShare cable that is appropriate for the multiparameter monitor that is PULSE BEEP VOLUME Set to pre-power down setting being connected. Check the Masimo web site at www.masimo.com for the latest list of available SatShare cables and validated instruments. TREND DISPLAY PARAMETERS...
  • Page 17: Philips Vuelink Setup

    Central Station monitor. Please see the RadNet Hardwired and Wireless Systems configure the Philips monitor. Please see the Philips Operator’s manual for complete Operator’s Manual for complete instructions. instructions. Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 18: Introduction

    The readings for SpO and pulse rate. NOTE: “—” may appear on the numeric display until the SpO and pulse rate readings have stabilized (approximately 10 seconds). Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 19: Successful Spo 2 Monitoring

    16. After monitoring is complete, remove the sensor from the patient and store or ■ dispose of the sensor according to governing rules. See the Directions for Use of DO NOT ATTEMPT TO REPROCESS, RECONDITION OR RECYCLE MASIMO SENSORS OR PATIENT CABLES AS THESE PROCESSES MAY DAMAGE THE the sensor.
  • Page 20: Numeric Display - Spo

    After performing the above, if the "Low Signal IQ" message is displayed frequently or continuously obtaining an arterial blood specimen for CO-Oximetry analysis may be considered to verify the oxygen saturation value. Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 21: Low Perfusion

    ■ If possible, ensure that the sensor is placed in a location with low ambient light. Although the Radical with integrated Masimo SET technology has significant immunity to ambient light, excessive ambient light may cause readings to be incorrect.
  • Page 22: Touch Key Control Button And Icons

    The lowest volume level will silence the pulse beep. DECREASE LOUDNESS Press the Decrease Loudness button to decrease the volume of the pulse beep. The lowest volume level will silence the pulse beep. LOUDNESS Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 23: Navigating The Main Menu

    Select the Next icon to scroll through the parameters. Once a NEXT parameter is highlighted, edit the parameter by pressing the Edit Parameter icon. 4-10 Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual 4-11...
  • Page 24: Alarms

    Polarity entered high alarm limit setting. Handheld Battery Deep Discharge SERVICE DS Battery Deep Discharge 4-12 Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual 4-13...
  • Page 25 In enabled, any values changed by the user will be returned to the default values after a power cycle . 4-14 Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual 4-15...
  • Page 26: Display

    Allows the user to set a custom default minimum low SpO 2 limit.The unit (meaning 10 saturation units) below the Low SpO alarm threshold. will return to this setting after a power cycle. 4-16 Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual 4-17...
  • Page 27: Output

    CAUTION: TO AVOID EXCESSIVE BATTERY DISCHARGING, DO NOT CONNECT ANY EQUIPMENT TO THE SERIAL PORT ON THE BACK PANEL UNLESS THE RADICAL IS CONNECTED TO THE AC MAINS POWER SUPPLY. 4-18 Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual 4-19...
  • Page 28: Service

    Selecting Trend from the Main, Display Menu. STATION BATTERY, MAKE SURE THAT THE DEVICE HAS BEEN REMOVED FROM SERVICE UNTIL FULL BATTERY CAPABILITY CAN BE RESTORED. 4-20 Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual 4-21...
  • Page 29: Navigating The Trend Display

    Radical display will be shown. CLEAR TREND Press the clear Trend Data button to clear the data stored in the trend memory. DATA 4-22 Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual 4-23...
  • Page 30: Trend Setup

    SpO PI and pulse rate measurement. Each trend record is printed on a single line, followed by a carriage return and line feed character. 4-24 Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual 4-25...
  • Page 31: Satshare Operation

    Normal or APOD sensitivity mode, increase the perfusion at the measurement site (by warming the patient or sensor site), or move the sensor to a site with better perfusion 4-26 Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual 4-27...
  • Page 32: Password Operation

    Audible and visual alarms for low battery. PRIORITY NOTE: High priority alarms indicate that immediate operator response is required. Low priority alarms indicate that operator awareness is required. 4-28 Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 33: System Messages

    The low SpO rate limit has been exceeded.. PR HIGH The high pulse rate limit has been exceeded. PR LOW The low pulse rate limit has been exceeded. Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 34: Troubleshooting

    Radical features normally accessed through the docking station, including SatShare, analog output and serial output will not be available to the user. Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 35: Radical Specifications

    55 VA Fuses: 1 Amp, Fast Acting, Metric, (5x20mm), 250V batteries Handheld: Type: NiMH Capacity: 8 hours Charging time: 3 hours Docking Station (RDS-1B, RDS-3B): Type: NiMH Capacity: 12 hours Charging time: 6 hours Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 36: Environmental

    3 The Radical with SatShare has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and Masimo’s simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 72 hours of trending at 2 second resolution, up to 30 days of trending at 10 second resolution , output to 100%.
  • Page 37: Serial Printer Setup

    Once serial communication is established, the Radical Analog 2 will automatically start printing the ASCII1 text data. WARNING: ALL EXTERNAL DEVICE CONNECTIONS TO THE RS-232 SERIAL PORT MUST BE IEC-60950 COMPLIANT. Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 38: Analog Output

    0.0V to 1.0V. A variance of ± 40 mV is acceptable. or sensors may cause improper Radical performance. Tissue damage can be caused by incorrect application or use of a Masimo sensor, for CALIBRATION example by wrapping the sensor too tightly. Inspect the sensor site as directed in the...
  • Page 39: Sensor Application Site

    ± 3% ± 3 bpm Bridge LNOP NeoPt < 1 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm Bridge Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 40: Lncs

    LNCS ADHESIVE SENSORS (LNCS sensors must be used in conjunction with LNC cables) Reusable patient cables of various lengths are available. Only use appropriate Masimo Weight Saturation Accuracy Pulse Rate Accuracy oximetry patient cables for SpO measurements.
  • Page 41: Introduction

    HARSH SOLVENTS, TO CLEAN THE OXIMETER. THESE SUBSTANCES ATTACK THE DEVICE’S MATERIALS AND DEVICE FAILURE CAN RESULT. Refer to Section 8, Cleaning and Reuse of Masimo Sensors for cleaning instructions of the sensor. Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 42: Battery Operation And Maintenance

    To properly discharge the battery pack, use the Battery Discharge function as described in Section 4, Service . Power Save “no” STANDALONE 6 hrs Backlight turned “on” Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 43: Replacing The Batteries

    Set the Radical to normal operating mode by selecting the Home Use parameter in the General Menu to “No”. Radical Signal Extraction Pulse Oximeter Operator’s Manual Radical Signal Extraction Pulse Oximeter Operator’s Manual...
  • Page 44 Reset the alarm limits again to the original settings. Nurse Call Test: Disconnect the patient cable or the Masimo SET Tester from the Radical and turn the Display Contrast Test: instrument on. Ensure that there are no audible alarms and that the audible alarms With the monitor turned on, select the Menu Access key and enter the Display menu.
  • Page 45: Service And Repair

    Verify that the green Handheld Battery Indicator LED is lit up. Please clean contaminated/dirty equipment before returning and make sure it is fully dry before packing the equipment. Call Masimo at 800-326-4890 and ask for Technical Support. When the Radical is fully charged the green Handheld Battery Indicator turns off.
  • Page 46: Exclusions

    MASIMO software and/or firmware and the documentation, and all copies thereof, remain at all times vested in MASIMO Corporation, licensor to MASIMO, and Masimo warrants to the initial purchaser that each new pulse oximeter will be free from they do not pass to Purchaser.
  • Page 47: Section 10 - Accessories

    11903 RADICAL USER MANUAL - SWEDISH 12114 RADICAL SERVICE MANUAL 30475 RADICAL HANDHELD CASE 1593 MASIMO TESTER 1795 MASIMO TESTER WITH CABLE Please visit our website, www.masimo.com, for updated information about accessories. Radical Signal Extraction Pulse Oximeter Operator’s Manual 10-1...
  • Page 48 Instruments and sensors containing Masimo SET technology are identified with the Masimo SET logo. Look for the Masimo SET designation on both the sensors and monitors to ensure accurate pulse oximetry when needed most. © 2007 Masimo Corporation. Masimo, Radical, Discrete Saturation Transform, DST, Satshare,SET, LNOP, LNCS and LNOPv are federally registered trademarks of Masimo Corporation.