Masimo Radical-7R Operator's Manual

Masimo Radical-7R Operator's Manual

Pulse co-oximeter with rainbow technology
Table of Contents

Advertisement

Quick Links

Pulse CO-Oximeter
with Rainbow
Technology
OPERATOR'S
MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Radical-7R and is the answer not in the manual?

Questions and answers

Summary of Contents for Masimo Radical-7R

  • Page 1 Pulse CO-Oximeter with Rainbow Technology OPERATOR’S MANUAL...
  • Page 2 The Radical-7R Pulse CO-Oximeter Operating Instructions provide the necessary information for proper operation of all models of the Radical-7R Pulse CO-Oximetry system. There may be information provided in this manual that is not relevant for your system. General knowledge of pulse oximetry and an understanding of the features and functions of the Radical-7R Pulse CO-Oximeter are prerequisites for its proper use.
  • Page 3 Variation in hemoglobin measurements may be profound and may be affected by sample type, body positioning, as well as other physiological conditions. As with most hemoglobin tests, Radical-7R test results should be scrutinized in light of a specific patient’s condition. Any results exhibiting inconsistency with the patient’s clinical status should be repeated and/or supplemented with additional test data.
  • Page 4 ■ Always remove the sensor from the patient and completely disconnect the patient from the Pulse CO-Oximeter before bathing the patient. Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 5 SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES (CONTINUED) ■ Additional information specifi c to Masimo sensors, including information about parameter/ measurement performance during motion and low perfusion, may be found in the sensor's Directions for Use (DFU). ■ Do not place the Pulse CO-Oximeter where the controls can be changed by the patient.
  • Page 6: Table Of Contents

    Radical-7R Pulse CO-Oximeter Handheld........2-2...
  • Page 7 Masimo Pulse CO-Oximetry Sensors........4-3...
  • Page 8 Masimo Acoustic Respiration Sensor ........
  • Page 9: About This Manual

    A b o u t T h i s M a n u a l This manual explains how to set up and use the Radical-7R Pulse CO-Oximeter containing Masimo Rainbow SET ® technology. Important safety information relating to general use of the Pulse CO-Oximeter appears before this introduction.
  • Page 10: Warnings, Cautions And Notes

    Sample of Caution: CAUTION: THIS IS A SAMPLE OF A CAUTION STATEMENT. A NOTE is provided when extra general information is applicable. Sample of Note: NOTE: This is a sample of a Note. Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 11: Product Description

    Pleth Variability Index (PVI ® ). It also provides graphical displays for plethysmographic waveform and Signal Identifi cation and Quality Indicator (Signal IQ ® ). The Radical-7R Pulse CO-Oximeter can be used to interface with a multiparameter patient monitor to provide Masimo SET SpO 2 and pulse rate information to that monitor for display.
  • Page 12: Indications For Use

    O v e r v i e w INDICATIONS FOR USE The Radical-7R Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa).
  • Page 13: Sphb General Description

    Hematocrit is the fraction of whole blood volume that consists of red blood cells. In nor- mal conditions, there is a linear relationship between hematocrit and the concentration of hemoglobin. An estimated hematocrit as a percentage may be derived by multiplying Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 14: Principle Of Operation

    (7 + wavelengths) 2. Detector Once the Radical-7R receives the signal from the sensor, it utilizes Masimo Rainbow SET signal extraction technology to calculate the patient’s functional oxygen saturation (SpO 2 (%)), blood levels of carboxyhemoglobin (SpCO (%)), methemoglobin (SpMet (%)), Total Hemoglobin concentration (SpHb (g/dl)) and pulse rate (PR (BPM)).
  • Page 15: Functional Saturation

    In an ambient temperature of 35º C the maximum skin surface temperature has been measured at less than 106º F (41º C), verifi ed by Masimo sensor skin temperature test procedure.
  • Page 16: Spco, Spmet And Sphb Measurements During Patient Motion

    O v e r v i e w SPCO, SPMET, AND SPHB MEASUREMENTS DURING PATIENT MOTION The Radical-7R displays measurements of SpCO, SpMet and SpHb during patient motion. However, because of the changes in the physiological parameters such as blood volume, arterial-venous coupling, etc., that occur during patient motion, the accuracy of such measurements may not be...
  • Page 17 [5] Gavriely N, Cugell DW. Airflow effects on amplitude and spectral content of normal breath sounds. J Appl Physiol 1996; 80: 5-13. [6] Gavrieli N, Palti Y, Alroy G. Spectral characteristics of normal breath sounds. J Appl Physiol 1981; 50: 307-314. Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 18: Masimo Rainbow Set Parallel Engines

    O v e r v i e w MASIMO RAINBOW SET PARALLEL ENGINES This fi gure is for conceptual purposes only. SpOC ® MASIMO SET DST 1-10 Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 19: Introduction

    ■ The Radical-7R is a fully featured Standalone Pulse CO-Oximeter. ■ The Radical-7R interfaces to the SpO 2 input module of multiparameter patient monitors to upgrade conventional pulse oximetry technology to Masimo SET technology. The Handheld portion of the Radical-7R contains the majority of the Pulse CO-Oximeter features.
  • Page 20: Radical-7R Pulse Co-Oximeter Handheld

    The Handheld Radical-7R Pulse CO-Oximeter provides most of the functionality of the Pulse CO-Oximeter. All user input and displays are controlled by this part of the Radical-7R Pulse CO-Oximeter system. The patient cable connects into the connector on the Handheld instrument.
  • Page 21 “Sensor Off” will appear at DISPLAY the top of the display screen. The display will show dashed lines during pulse search. The total hemoglobin is calculated and the display is updated at a frequency of once per second Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 22 Connect a patient cable or a direct cable sensor into the Handheld PATIENT CABLE Radical-7R by plugging the cable into the Patient Cable Connector. Use CONNECTOR only Masimo compatible sensors and cables with this Pulse CO-Oximeter.
  • Page 23 The Perfusion Index indicates numerically the percentage of INDEX pulsatile signal to non-pulsatile signal (pulse strength). The sensitivity icon is shown on the Radical-7R display to indicate if the Radical 7R is set to operate in Normal, Maximum (MAX) or APOD SENSITIVITY Adapative Probe Off Detection (APOD) mode.
  • Page 24: Handheld Back Panel

    See the Directions for Use of the Pole Clamp accessory for ACCESSORY attachment instructions. HOLDER The Radical-7R Handheld is powered by a NiMH battery BATTERY PACK located in this compartment. For battery care and replacement please see Section 9, Replacing the Batteries.
  • Page 25: Radical-7R Pulse Co-Oximeter Standalone

    R a d i c a l - 7 R P u l s e C O - O x i m e t e r S t a n d a l o n e When the Radical-7R Pulse CO-Oximeter Handheld is placed into the Docking Station, the Radical-7R Pulse CO-Oximeter becomes a full-featured standalone instrument. The Radical-7R Pulse CO-Oximeter Standalone acts as a battery charger for the Handheld instrument and has AC power connection capabilities.
  • Page 26: Standalone Front Panel

    S y s t e m D e s c r i p t i o n STANDALONE FRONT PANEL The following figure and corresponding text review the features of the Radical-7R Standalone instrument. 10.0 SpHb g/dl SpOC ml/dl DOCKING...
  • Page 27: Standalone Back Panel

    STANDALONE BACK PANEL Use the Serial Output Connector with a ferrite bead installed to connect a serial instrument, including a serial printer, a monitoring system or PC to the Radical-7R Pulse SERIAL OUTPUT CO-Oximeter. The data is provided in standard RS-232C CONNECTOR format.
  • Page 28: Symbols

    S y s t e m D e s c r i p t i o n SYMBOLS The following symbols may be found on the Radical-7R Pulse CO-Oximeter, Docking Station or packaging and are defined below: NOTE: Some of the interfaces and symbols are not available on all versions of the Docking Station.
  • Page 29: Radical-7R Monitor Interface

    R a d i c a l - 7 R M o n i t o r I n t e r f a c e In addition to being a full-featured Handheld and Standalone Pulse CO-Oximeter, the Radical-7R Pulse CO-Oximeter’s unique SatShare interface links the Radical-7R Pulse CO-Oximeter to most existing multiparameter patient monitors through the pulse oximetry patient cable or SpO input connector.
  • Page 30: Introduction

    S e t u p I n t r o d u c t i o n Before the Radical-7R Pulse CO-Oximeter can be used in a clinical setting, it needs to be inspected, properly setup and the batteries need to be fully charged.
  • Page 31: Initial Battery Charging

    Place the Docking Station on a stable hard flat surface near the patient. Always place the Radical-7R Pulse CO-Oximeter instrument on a dry surface. Maintain a minimum of 3 cm (1 inch) free space around the Radical-7R Pulse CO-Oximeter Standalone instrument. Make sure that the Radical-7R speaker is not covered to avoid a muffled alarm sound.
  • Page 32: Satshare Setup

    S e t u p S a t S h a r e S e t u p The Radical-7R Pulse CO-Oximeter has been proven to be accurate during patient motion and low perfusion conditions. Saturation and pulse rate values from the Radical-7R Pulse CO-Oximeter may be displayed on a multiparameter monitor through the SatShare feature.
  • Page 33 AC power, set the Power Save parameter in the General menu to “No” and refer to Section 4 Operation. Please note that if the Radical-7R Pulse CO-Oximeter is used in this mode, the length of time the Radical-7R Pulse CO-Oximeter can operate on battery power will be significantly diminished.
  • Page 34 S e t u p THE RADICAL-7R REVERTS TO INTERFACE ALARM SpO 2 /BPM "YES" DURING POWER INTERRUPTIONS OR WHEN THE SATSHARE CONNECTION IS LOST OR THE INSTRUMENT BECOMES SEPARATED FROM THE DOCKING STATION. THIS ENSURES THAT THE RADICAL- 7R PROVIDES AUDIBLE ALARMS FOR SpO 2 AND PULSE RATE WHEN THE CONNECTION TO THE INTERFACED SYSTEM BECOMES COMPROMISED.
  • Page 35: Introduction

    NOTE: “- - -” initially shows in the numeric display fields for all the parameters/measurements when the Radical-7R is turned on. As the system starts monitoring, the numeric display fields update (refresh). The numeric display fields for the parameters/measurements begin to show numbers during the refresh cycles even though the numbers have not stabilized;...
  • Page 36: Successful Monitoring

    Place the sensor on a site that has unrestricted blood flow. ■ Do not secure a sensor with tape. ■ Do not select a site near potential electrical interference (electrosurgical unit, for example). ■ Read the sensor Directions for Use for proper sensor application. Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 37: Masimo Pulse Co-Oximetry Sensors

    ■ Turn the power off and on at the instrument. If a Masimo Rainbow Adhesive Sensor is being used and “SpO2 Only Mode” appears on the display screen, perform one of the following steps to to obtain Rainbow values: ■...
  • Page 38: Numeric Display - Spo

    Motion artifact. NUMERIC DISPLAY - PULSE RATE The Pulse Rate displayed on the Radical-7R Pulse CO-Oximeter may differ slightly from the heart rate displayed on ECG monitors due to differences in averaging times. There may also be a discrepancy between cardiac electrical activity and peripheral arterial pulsation. Significant differences may indicate a problem with the signal quality due to physiological changes in the patient or one of the instruments or application of the sensor or patient cable.
  • Page 39: Numeric Display - Sphb

    Inaccurate calculations may be caused by: ■ Intravascular dyes such as indocyanine green or methylene blue. ■ Low arterial perfusion. ■ Low arterial oxygen saturation levels. ■ Elevated Total Bilirubin levels. ■ Motion artifact. Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 40: Signal Indication And Quality Indicator (Siq)

    When the SpHb signal quality is very low, the accuracy of the SpHb measurement may be compromised. A Low SpHb SIQ message is displayed in the message area on the Radical-7R Pulse CO-Oximeter display, and the parameter/measurement value will display dashes (“---”) instead of a number value for SpOC and SpHct.
  • Page 41: Signal Indication And Quality Indicator (Siqa)

    ■ Check the Acoustic Respiration sensor and ensure proper sensor application. The Acoustic Respiration sensor must be well secured to the site for the Radical-7R to maintain accurate readings. Refer to the Acoustic Respiration sensor's Directions for Use for proper sensor placement.
  • Page 42: Numeric Display - (Pi)

    PVI is displayed as a percentage (0-100%). LOW PERFUSION The Radical-7R Pulse CO-Oximeter displays a “Low PI” message when there are very low amplitude arterial pulsations. It has been suggested that at extremely low perfusion levels, pulse oximeters can measure peripheral saturation, which may differ from central arterial saturation 3 .
  • Page 43: Sensitivity

    PATIENT’S VITAL SIGNS BY ALTERNATE MEANS AND THEN CHECK THE PULSE CO-OXIMETER FOR PROPER FUNCTIONING. SENSITIVITY Three sensitivity levels enable a clinician to tailor the response of the Radical-7R to the needs of the particular patient situation. They are as follows: ■...
  • Page 44: Touch Key Control Button And Icons

    Touch Key Control Buttons and Icons On the Radical-7R Pulse CO-Oximeter display, four icons are shown on the right side of the LCD display. These icons can be selected either by using the touch key control buttons or the touchscreen.
  • Page 45: Navigating The Main Menu

    Select the Accept icon to save the changes. ACCEPT Select the Previous icon to increase or toggle the parameter/ PREVIOUS measurement settings. Select the Next icon to decrease or toggle the parameter/ NEXT measurement settings. Radical-7R Pulse CO-Oximeter Operator’s Manual 4-11...
  • Page 46: Menu Tree

    O p e r a t i o n MENU TREE MAIN MENU ALARM LIMITS MENU 4-12 Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 47 O p e r a t i o n GENERAL MENU ABOUT MENU CONFIGURATION MENU Radical-7R Pulse CO-Oximeter Operator’s Manual 4-13...
  • Page 48 O p e r a t i o n AVERAGING MENU OUTPUT MENU FACTORY DEFAULTS MENU 4-14 Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 49: Alarms

    NOTE: The low alarm limit must be set below the high alarm setting. If the high alarm limit is set below the low alarm limit, the low alarm limit will automatically adjust to the next setting below the newly entered high alarm limit setting. Radical-7R Pulse CO-Oximeter Operator’s Manual 4-15...
  • Page 50 This menu allows the users to set an audible saturation delay. The delay DELAY can be set to either 0, 5,10 or 15 seconds. The delay setting only affects saturation alarms indications. 4-16 Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 51: General

    NOTE: The hemorheologic profi le of arterial and venous blood samples can vary 1 . To accommodate this difference, the SpHb Cal Radical-7R provides the option of displaying a SpHb parameter that is based on either Arterial or Venous SpHb laboratory blood sample draw.
  • Page 52: Output

    Closed contacts. See Section 7, Nurse Call specifi cations. CAUTION: TO AVOID EXCESSIVE BATTERY DISCHARGING, DO NOT CONNECT ANY EQUIPMENT TO THE SERIAL PORT ON THE BACK PANEL UNLESS THE RADICAL-7R IS CONNECTED TO THE AC POWER FROM THE WALL OUTLET. 4-18...
  • Page 53: Satshare Operation

    S a t S h a r e O p e r a t i o n When the SatShare cable is connected to the Radical-7R Pulse CO-Oximeter and to a multiparameter patient monitor, the Radical-7R Pulse CO-Oximeter automatically starts to operate in the SatShare mode.
  • Page 54 SatShare interface if the perfusion index drops below 0.1% while the sensitivity is set to Max sensitivity. To enable the SatShare interface again, set the Radical-7R Pulse CO-Oximeter to the Normal or APOD sensitivity mode, increase the perfusion at the measurement site (by warming the patient or sensor site), or move the sensor to a site with better perfusion.
  • Page 55: Alarm Identification

    A l a r m I d e n t i f i c a t i o n The Radical-7R visually and audibly indicates alarm conditions that the system detects. Audible alarms may be silenced, without affecting the operation of visual alarms.
  • Page 56: System Messages

    Correct the correct adhesive por- REPLACE unrecognized. (Applies to Rainbow tion of the sensor. ADHESIVE ReSposable CO-Oximeter Sensor Systems, or ReSposable Pulse Oximeter Sensor Systems only) Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 57 "SpO 2 Only Mode" message for Use instructions, Section occurs during an unsucessful SPO 2 ONLY MODE 4, Sucessful Monitoring, and sensor calibration/pulse search Section 8, Selecting a Masimo routine, or during monitoring. SET Sensor. Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 58 Not a proper cable. Section 8. May be an incorrect cable, or a defective NO AC CABLE AC Cable not detected. cable. Disconnect and reconnect the cable. Refer to instructions for the cable being used. Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 59 Disconnect and reconnect into the connector. sensor. Refer to the instructions for the sensor being used. INCOMPATIBLE AC Replace with a proper Masimo sensor. Refer Not a proper Masimo sensor. SENSOR to Section 8. Pulse CO-Oximeter Inoperative or faulty AC sensor; Replace...
  • Page 60: Troubleshooting

    T r o u b l e s h o o t i n g T r o u b l e s h o o t i n g The following chart describes what to do if the Radical-7R Pulse CO-Oximeter system does not operate properly or fails.
  • Page 61 OF DOCKING STATION Match handheld to docking station with handheld and docking CONTINUOUSLY FLASH compatible software versions. station. BATTERY RUN-TIME IS Use Battery Discharge function as Battery Memory effects. SIGNIFICANTLY REDUCED described in Section 4, Service. Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 62 SpCO/SpMet/SpHb READING Excessive ambient or Shield the sensor from excessive or strobing light. strobing light. Minimize or eliminate motion at the Excessive motion. monitoring site. Also, see Section 4, Successful Monitoring for additional information. Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 63: Radical-7R Specifications

    1% - 15% ± 1% Total Hemoglobin accuracy (SpHb g/dl) 6 Adults, Pediatrics 8 - 17 g/dl ±1 g/dL Respiratory Rate Accuracy (RRa, breaths per minute) 7 4 to 70 ± 1 breath per minute Adults Resolution Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 64: Electrical

    4.11 lbs (1.86 kg) Standalone (RDS-1, RDS-2, RDS-3): 3.8 lbs. (1.73 kg) Standalone (RDS-1B): 5.4 lbs. (2.45 kg) Mode 2, 4, 8,10, 12, 14 or 16 seconds 10 Averaging mode: Normal and Maximum 11 and APOD Sensitivity: Alarms Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 65 0.9% SpMet. Contact Masimo for testing specifications. 2 The Masimo sensors have been validated for no motion accuracy in human blood studies on healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range of 70-100% SpO 2 against a laboratory CO-Oximeter and ECG monitor.
  • Page 66: Analog Output/Nurse Call Specifications

    68% of the population. 5 Masimo sensors have been validated for pulse rate accuracy for the range of 25-240 bpm in bench top testing against a Biotek Index 2 simulator. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
  • Page 67: Nurse Call

    0.0V to 1.0V. A variance of ± 40 mV is acceptable. NURSE CALL The nurse call feature is available when Radical-7R Pulse CO-Oximeter is operating in its standalone configuration. The nurse call feature on the Radical-7R Pulse CO-Oximeter is based on the relay closing or opening depending on alarm, Low Signal IQ events or both.
  • Page 68: Introduction

    To prevent interference from ambient light, ensure that the sensor is properly applied, and cover the sensor site with a Masimo Optical Shield, if required. Failure to take this precaution in excessive ambient light conditions may result in inaccurate measurements.
  • Page 69: Masimo Sensors

    The following sensors are compatible for use with the Radical-7R Pulse CO-Oximeter. See the sensor directions for use for full sensor specifications, including accuracy. MASIMO ACOUSTIC RESPIRATION SENSOR The Acoustic Respiration sensor must be used with the Radical-7R Pulse CO-Oximeter to enable monitoring of respiration rate (RRa). Acoustic Respiration Sensor The Acoustic Respiration sensor must be used in conjunction with the Acoustic Respiration PC and Dual Rainbow Cable.
  • Page 70: Rainbow Direct Connect Sensors

    • S2-20a with S2-20r Red Direct Connect Sensors Masimo Red sensors can be used with the Radical-7R to enable measurement of SpO 2 and pulse rate only. Red sensors will only function with Pulse CO-Oximeter instruments equipped with Masimo Rainbow SET technology. Red Direct Connect sensors connect to the instrument directly.
  • Page 71: Lnop ® Reusable Sensors

    LNOP sensors must be used in conjunction with Red PC cables. • Newborn Infant/Pediatric • Newborn Neonatal • Trauma • Blue ® M-LNCS™/LNCS Reusable Sensors LNCS sensors must be used in conjunction with LNC cables. • DCI • DCIP • YI Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 72: M-Lncs™/Lncs ® Adhesive Sensors

    LNCS sensors must be used in conjunction with Red LNC cables. • Adtx • Pdtx • Inf • Neo • NeoPt • NeoPt-500 ® M-LNCS™/LNCS Specialty Sensors LNCS specialty sensors must be used in conjunction with Red LNC cables. • Newborn Infant/Pediatric Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 73: Sensor Accuracy

    See Sensor Directions for Use (DFU) for sensor accuracy specifications for: SpO 2 , SpMet, SpCO, SpHb, pulse rate and respiration rate. CLEANING AND REUSE OF MASIMO REUSABLE SENSORS AND CABLES Reusable sensors and patient cables can be cleaned per the following procedure: Remove the sensor from the patient.
  • Page 74: Introduction

    DO NOT USE PETROLEUM-BASED OR ACETONE SOLUTIONS, OR OTHER HARSH SOLVENTS, TO CLEAN THE PULSE CO-OXIMETER. THESE SUBSTANCES ERODE THE INSTRUMENT’S MATERIALS AND INSTRUMENT FAILURE CAN RESULT. Refer to Section 8, Cleaning and Reuse of Masimo Sensors for cleaning instructions of the sensor. Radical-7R Pulse CO-Oximeter Operator’s Manual...
  • Page 75: Battery Operation And Maintenance

    IT IS NOT IN USE TO ENSURE THAT THE BATTERY REMAINS FULLY CHARGED. The following tables outline the estimated run times of the battery powered Radical-7R Pulse CO-Oximeter. The time estimates are based on a Radical-7R Pulse CO-Oximeter with fully charged batteries.
  • Page 76: Replacing The Batteries

    Radical-7R Pulse CO-Oximeter configured to only include the Handheld battery (standard configuration); the Docking Station battery is excluded. NOTE: For this configuration, it is advisable to operate only the Radical-7R Handheld instrument when running on battery power. Although it is possible to operate the entire Standalone...
  • Page 77: Performance Verification

    Before performing the following tests place the Radical-7R Pulse CO-Oximeter Handheld into the Docking Station, connect the Radical-7R to AC power and fully charge the Radical-7R Pulse CO-Oximeter Handheld battery. Also disconnect any patient cables or pulse oximetry probes,...
  • Page 78 Masimo SET Tester's Directions for Use for further instructions. Nurse Call Test: Disconnect the Red patient cable or the Masimo SET Tester from the Radical-7R and turn the instrument on. Ensure that there are no audible alarms and that the audible alarms are not silenced.
  • Page 79: Service And Repair

    Please clean contaminated/dirty equipment before returning. Make sure the equipment is fully dry before packing. Call Masimo at 800-326-4890 and ask for Technical Support. Ask for an RMA number. Package the equipment securely – in the original shipping container if possible – and enclose or include the following information and items: ■...
  • Page 80 S e r v i c e / M a i n t e n a n c e Return Radical-7R Pulse CO-Oximeter to the following shipping address: For USA, Canada & Asia Pacific: For Europe: All other locations: Masimo Corporation Contact your Masimo International Sàrl...
  • Page 81: Sales & End-User License Agreement

    W a r r a n t y Masimo warrants to the initial Purchaser for a period of one (1) year from the date of purchase that: (i) each new Product and the Software media as delivered are free from defects in workmanship or materials, and (ii) the Product and Software will perform substantially as labeled in the directions for use.
  • Page 82: End-User License Agreement

    Ownership of Software: The Software is licensed not sold; all rights and interests in the Software and all copies thereof remain at all times vested in Masimo, and do not pass to Purchaser. Any references in this Agreement to the purchase or sale of the Software shall be deemed the purchase or sale of a Software License as set forth herein.
  • Page 83: Sensors Licensed For Monitoring Use Only

    Masimo. There is no license, implied or otherwise, that would allow use of licensed sensors beyond their intended duration. After use of sensors through their designated duration, there is no further license granted by Masimo to use the sensors and the sensors must be discarded or returned to Masimo for reprocessing.
  • Page 84: Part Numbers

    RADICAL-7R POLE CLAMP 2350 RADICAL-7R HANDHELD LOCK 2351 RADICAL-7R HANDHELD LOCK KEY 2368 MASIMO SET TESTER - 20 PIN 1584 RADICAL-7R POWER CORD LOCK, 5/PACK Please visit our website, www.masimo.com, for updated information about Masimo products. Radical-7R Pulse CO-Oximeter Operator’s Manual 10-1...
  • Page 85 Instruments and sensors containing Masimo Rainbow SET technology are identified with the Masimo Rainbow SET logo.