Root NIBPT
Procedure
1.
Connect the manometer, 500mL bottle or cuff wrapped around solid object, and
hand bulb to the NIBP module using "T" adapters and connection tubing.
2.
Power ON the manometer.
•
Verify the manometer has been zeroed.
•
Set the unit of measurement to mmHg.
•
Verify the manometer has been calibrated.
3.
Power ON the Root.
4.
On the Root home screen, press the Main Menu icon
Press NIBP
5.
Press CALIBRATION
6.
•
For the password, enter 4258.
•
Press the Enter key.
Press CALIBRATION TEST
7.
Press Test.
8.
www.masimo.com
.
.
.
151
Chapter 16: Service and Maintenance
.