Root NIBPT
Access Control
Access Control contains configurable options and settings that require a password.
To enter Access Control
1.
Press the
2.
When the numeric screen displays, enter the following numbers: 6 2 7 4
Asterisks (****) will be displayed.
To undo an entry, press Backspace.
Press Enter to access the password protected screen.
3.
Note: The password will have to be entered every time this screen is accessed.
Option
Power On Profile Sets the profile used when
All Mute
Enabled
Lock Alarm
Volume
Optical Sensor
Off Alarm Delay
www.masimo.com
key.
Description
the device is powered on.
Enables or disables
parameter Alarm Silence
menu option. See Sounds
on page 63.
Sets the lowest alarm
volume level.
Enables or disables alarm
delay when Optical Sensor
is off.
70
Chapter 3: Operation
Factory
Configurable
Default
Settings
Setting
Adult, Adult
Previous
Modified, Neonatal,
Profile
Pediatric, or
Previous Profile
Off
On or Off
Off
3, 4, or Off
0, 5, 10, 15, 30, or
0 Sec.
60 seconds
Masimo
Need help?
Do you have a question about the Root and is the answer not in the manual?