Chapter 15: Specifications

This chapter contains specifications of Root.
For information on the specifications of Radical-7, Radius-7, MOC-9 modules, and standalone
devices, see Directions for Use or Operator's Manuals for these devices.

Measurement Accuracy

NIBP
Pressure Transducer
Between 0 mmHg and 300 mmHg ±3mmHg
Optional Integrated Temperature*
Temperature
Between 80°F and 110°F
(26.7°C and 43.3°C)
Temperature
Measurement Site/Mode
Oral
Rectal
Pediatric Axillary
Adult Axillary
*Test Report #:2010036 Test Report of Clinical Investigation to demonstrate that the
SureTemp® Plus thermometer meets the essential performance of clinical thermometers for
body temperature measurement as described in ISO 80601-2-56:2009
**Applicable to Continuous Mode only. See Operation - Temperature on page 89.
www.masimo.com
All patient populations
Number of
Clinical
Subjects
Bias (°C)
106
0.01
105
-0.12
117
-0.03
105
0.13
133
±0.2°F
(0.11°C)**
Limits of
Clinical
Agreement
Repeatability
0.63
0.14
0.59
0.29
0.56
0.14
0.43
0.14