Root NIBPT
Systolic/Diastolic (SYS/DIA)
From the Systolic/Diastolic screen, touch Alarms, and then change any of the following
options:
Options
Description
The High Limit is upper
Systolic
threshold that triggers an
High Limit
alarm.
Low Limit is the lower
Systolic Low
threshold that triggers an
Limit
alarm.
The High Limit is upper
Diastolic
threshold that triggers an
High Limit
alarm.
Low Limit is the lower
Diastolic
threshold that triggers an
Low Limit
alarm.
www.masimo.com
Factory
Alarm
Default
Priority
Settings
Medium
220
Medium
75
Medium
110
Medium
35
59
Chapter 3: Operation
Configurable
Options
42-259 in steps of 1,
or Off
When set to Off,
alarm is disabled
41-258 in steps of 1,
or Off
When set to Off,
alarm is disabled
22-199 in steps of 1,
or Off
When set to Off,
alarm is disabled
21-198 in steps of 1,
or Off
When set to Off,
alarm is disabled