Root NIBPT
Mean Arterial Pressure (MAP)
From the Mean Arterial Pressure screen, touch Alarms, and then change any of the following
options:
Options Description
The High Limit is upper
High
threshold that triggers an
Limit
alarm.
Low Limit is the lower
Low
threshold that triggers an
Limit
alarm.
Intervals for NIBP
From the NIBP screen, select Intervals, then select a Set Mode: Automatic, Stat, or Schedule.
See the table below for factory default and user configurable settings:
Set Mode
Description
Automatic interval
measurement mode will
take blood pressure
Automatic
measurements once every
defined interval. See Set
Mode: Automatic on page
97.
www.masimo.com
Factory
Alarm
Default
Priority
Settings
Medium
120
Medium
50
Factory
Configuration
Default
Settings
Interval
15 min
61
Chapter 3: Operation
Configurable
Options
28-219 in steps of 1,
or Off
When set to Off,
alarm is disabled
27-218 in steps of 1,
or Off
When set to Off,
alarm is disabled
User Configurable
Settings
2 min, 3 min, 4 min, 5
min, 10 min, 15 min, 30
min, 60 min, 90 min
Masimo