LED on
LED on
power sup
battery
ply
charger
14.2.3 Error messages while establishing a connection with the cockpit app
Error message
Device not found
Component was connec
ted
to
another
device.
Establish connection?
14.2.4 Status signals
Battery charger is connected
LED on
LED on
power sup
battery
ply
charger
Battery charger disconnected
Beep sig
Vibration
nal
signal
1x short
1x short
3x short
Battery charge level
Battery
charger
Battery is being charged, battery charge level is less than 50%
Battery is being charged, battery charge level is over 50%
Battery is fully charged (or connection with product is interrupted).
Take note of the beep signals to differentiate between these states.
When the battery charger is connected or disconnected, a self-test is conducted and confirmed by
a single beep and vibration signal.
C-Leg 3C98-3/3C88-3
Error
Battery is fully charged (or connection
with product is interrupted).
Cause
The connection could not be
established as no component
with the entered serial num
ber could be found.
The component was connec
ted to another device/remote
control
A current connection to the
prosthesis was interrupted
Event
Power supply and battery charger operational
Event
Self-test completed successfully. Product is operational.
Maintenance note
Conduct the self-test again by connecting/disconnecting the battery charger. If the
beep signal sounds again, product maintenance should be carried out by an author
ised Ottobock Service Centre.
The product can be used without restrictions. However, vibration signals may not be
generated.
Resolution
If the connection to the product is inter
rupted, an authorised Ottobock Service
Centre must inspect the product, battery
charger and power supply.
Correction
Compare the serial number entered with that of the
component, then try to establish the connection
again.
To disconnect the original connection, tap the "OK"
button.
If the original connection is not to be disconnected,
tap the "Cancel" button.
Check the following points:
•
Distance from the prosthesis to the device
•
Battery charge level of the prosthesis
•
Bluetooth on the prosthesis switched on? (see
Page 69)
•
Prosthesis switched on? (see Page 69)
•
If multiple prostheses were stored in the remote
control, was the correct prosthesis selected?
Appendices
Ottobock | 75