Additional operating states (modes)

9.6.3 Status information on the display of the remote control (optional accessory)

Menu option
Trip:1747
Step:1747
Batt.:68
Bluetooth: On
10 Additional operating states (modes)

10.1 Empty battery mode

Beeps and vibration signals are emitted if the available battery charge level is 0% (see Page 74). During this time,
damping settings are set to their safety mode values. The prosthesis is then switched off. You can switch back to
basic mode (mode 1) from empty battery mode by charging the product.

10.2 Mode for charging the prosthesis

The product is non-functional during charging.
The product is set to safety mode damping. This may be low or high depending on the setting in the adjustment
software.

10.3 Safety mode

The product automatically switches to safety mode if a critical system fault occurs (e.g. failure of a sensor signal).
Safety mode remains in effect until the error has been rectified.
Default damping values are activated in safety mode. This makes limited walking possible for the user even though
the system is not active.
The switch to safety mode is indicated by beeps and vibration signals immediately prior to switching (see Page 74).
Safety mode can be disabled by connecting then disconnecting the battery charger. If the product switches into
safety mode again, this means a permanent error exists. The product must be inspected by an authorised Ottobock
Service Centre.

10.4 Overheating mode

When the hydraulic unit overheats due to uninterrupted, increased activity (e.g. extended walking downhill), damp­
ing is increased along with the rising temperature in order to counteract the overheating. When the hydraulic unit
cools down, the product switches back to the damping settings that existed before the overheating mode.
Overheating mode is indicated by a long vibration every 5 seconds.
When overheating mode is active, it is not possible to switch to a MyMode, display the current battery charge level
or change the prosthesis settings. In addition, the resistance in the flexion direction is not reduced when the sitting
function is active.
Overheating mode is not activated in the MyModes.

11 Cleaning

1) Clean the product with a damp cloth and mild soap (e.g. Ottobock 453H10=1 Derma Clean) when needed.
2) Dry the product with a lint-free cloth and allow it to air dry fully.

12 Maintenance

INFORMATION
This component was tested for three million load cycles in accordance with ISO 10328.
Depending on the activity level, this corresponds to a service life of three to five years.
The duration of use can be individually extended depending on the intensity of use by performing regular service
inspections.
70 | Ottobock
Description
Daily step counter
Total step counter
Current prosthesis charge level, as a
percentage
Switch Bluetooth function of the
prosthesis on or off
Possible actions
Reset the counter by using the  key
to confirm the menu option
For informational purposes only
For informational purposes only
The Bluetooth function of the pros­
thesis can be turned on or off (see
Page 69) by confirming the menu
item with the  key.
C-Leg 3C98-3/3C88-3

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

C-leg 3c88-3

Table of Contents