Use

9.3.1 Changing the prosthesis setting using the cockpit app

Product
MyModes
Functions
Settings
Status
General
Manage components
Imprint/Info

9.3.2 Overview of adjustment parameters in basic mode

The parameters in basic mode describe the dynamic behaviour of the prosthesis in a normal gait cycle. These para­
meters act as basic settings for automatically adjusting the damping behaviour to the current motion situation (e.g.
ramps, slow walking speed, etc.).
The stance function and/or the sitting function can also be activated/deactivated. Further information on the stance
function (see Page 63). Further information on the sitting function (see Page 64).
The following parameters can be modified:
Parameter
Resistance
Stance function
Sitting function
Volume

9.3.3 Overview of adjustment parameters in MyModes

The parameters in the MyModes describe the static behaviour of the prosthesis for a specific motion pattern such
as cross-country skiing. Damping behaviour is not automatically controlled and adjusted in MyModes.
The following parameters can be modified in MyModes:
Parameter
Gain
Volume
68 | Ottobock
1) Once the component is connected and in the desired mode, tap the
 symbol in the main menu.
→ The navigation menu opens.
2) Tap the menu entry "Settings".
change
→ A list appears with the parameters for the currently selected mode.
3) Change the setting of the desired parameter by tapping the "<", ">"
2. Golf
symbols.
INFORMATION: The prosthetist's setting is marked and, after the
setting has been changed, can be restored by tapping the "Stand­
ard" button.
Adjustment soft­
ware range
120 to 180
0 to 4
Adjustment soft­
ware range
0 to 100
0 — 4
App/remote
Meaning
control adjust­
ment range
+/- 10
Resistance against flexion motion e.g. when
walking down stairs or when sitting down
0 - deactivated
Activation/deactivation of the stance function.
1 - activated
This function needs to be enabled in the
adjustment software.
0 - deactivated
Activation/deactivation of the sitting function.
1 - activated
This function needs to be enabled in the
adjustment software.
0 to 4
Volume of beep signal for confirmation tones.
App/remote
Meaning
control adjust­
ment range
+/- 10
Value for speed at which flexion resistance
increases as the knee angle increases
0 — 4
Volume of beep signal for confirmation tones.
C-Leg 3C98-3/3C88-3