Cockpit App; System Requirements - Otto Bock C-Leg 3C98-3 Instructions For Use Manual

Qualified personnel
Hide thumbs Also See for C-Leg 3C98-3:
Table of Contents

Advertisement

Available languages

Available languages

Cockpit app

Verification at maximum flexion
Verification at maximum extension
8 Cockpit app
The cockpit app enables switching from basic mode into the pre-configured MyModes. In
addition, information about the product (step counter, battery charge level, etc.) can be
called up.
The behaviour of the product can be changed to a certain extent on a day-to-day basis
using the app (e.g. while becoming accustomed to the product). The adjustment software
can be used to trace the change at the patient's next appointment.
INFORMATION
The cockpit app can be downloaded free of charge from the respective online store. For more information, please
visit the website: http://www.ottobock-group.com/cockpitapp. You can also read the QR code on the supplied
Bluetooth PIN card.

8.1 System Requirements

The functioning of the cockpit app is assured on devices with the Android operating system version 4.0.3 or higher.
The functionality was tested on the following devices:
• Samsung
Galaxy S5, Galaxy S4, Galaxy S4 mini, Galaxy SIII, Galaxy SIII mini, Galaxy Note II, Galaxy
Fame
• Sony
Xperia Z, Xperia Z3, Xperia J, Xperia SP
• HTC
One, One mini
• LG
Optimus L9, Optimus F5, Optimus F6, Optimus G, Optimus G2
• Huawei
Ascend P6, Ascend G500
• Motorola
Droid Razr Maxx, Moto X, Nexus 6
60 | Ottobock
If the distance between the socket and hydraulics is not sufficient, the hydraulics
may be damaged. Check the distance as follows:
1) Bring the knee joint with socket to maximum flexion.
2) Check the available distance between the hydraulics and socket. It must be at
least 3 mm.
If the distance between the socket and electronics is not sufficient, the electronics
may be damaged. Check the distance as follows:
1) Bring the knee joint with socket to maximum extension.
2) Check the available distance between the electronics and socket. It must be at
least 5 mm.
For fittings with the C-Leg protector, the distance must be at least 10 mm.
INFORMATION: If the knee extender is used to verify the distance, note that
this is only permissible if the flexion stops are installed.