14.2.2 Warnings/error signals

Error during use
Beep signal
10x long
30x long
Error while charging the product
LED on
LED on
power sup­
battery
ply
charger
74 | Ottobock
Vibration signal
1x long at interval of
approx. 5 seconds
3x long
5x long
10x long
1x long, 1x short repeated
every 2.6 seconds
Error
Country-specific plug adapter not fully
engaged on power supply
Non-functional socket
Defective power supply
No connection between battery charger
and power supply
Defective battery charger
Battery is fully charged (or connection
with product is interrupted).
Event
Overheated hydraulic unit Reduce activity.
Battery
charge
level
under 25%
Battery
charge
level
under 15%
Battery charge level 0%
After the beep and vibra­
tion signal, the product
switches to empty battery
mode and then switches
off.
Severe error
e.g. failure of valve drives
Possibly no switching into
safety mode.
Resolution
Check whether the country-specific plug
adapter is fully engaged on the power
supply.
Check socket with another electric appli­
ance.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Check whether the charging cable plug is
fully engaged on the battery charger.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Take note of the beep signals to differenti­
ate between these states.
When the battery charger is connected or
disconnected, a self-test is conducted and
confirmed by a single beep and vibration
signal.
The battery is fully charged if this beep
signal is heard.
If no beep signal is heard, the connection
to the product is interrupted.
Required action
Charge battery soon.
Charge battery promptly, since
the product is switched off
after the next warning signal.
Charge the battery.
Restricted walking possible.
Attempt to reset this error by
connecting/disconnecting the
battery charger. The battery
charger must be connected for
at least 5 seconds before it is
disconnected.
If the error persists, it is no
longer permissible to use the
product. The product must be
inspected by an authorised
Ottobock Service Centre.