Configuration and Use Manual
MMI-20039472
10 Complete the configuration
10.1 Test or tune the system using sensor simulation
Display
Menu → Startup Tasks → Commissioning Tools → Sensor Simulation
ProLink III
Device Tools → Diagnostics → Sensor Simulation
Field Communicator
Service Tools → Simulate → Simulate Sensor
Use sensor simulation to test the system's response to a variety of process conditions, including boundary
conditions, problem conditions, or alert conditions, or to tune the loop.
Restriction
Sensor simulation is available only on flowmeters with the enhanced core processor.
Prerequisites
Before enabling sensor simulation, ensure that your process can tolerate the effects of the simulated process
values.
Procedure
1. Enable sensor simulation.
2. For mass flow, set Wave Form as desired and enter the required values.
Option
Fixed
Sawtooth
Sine
3. For density, set Wave Form as desired and enter the required values.
Option
Fixed
Sawtooth
Sine
Configuration and Use Manual
Required values
Fixed Value
Period
Minimum
Maximum
Period
Minimum
Maximum
Required values
Fixed Value
Period
Minimum
Maximum
Period
Minimum
Complete the configuration
June 2019
155