Emerson Micro Motion 5700 Configuration And Use Manual page 79

Transmitters with intrinsically safe outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configuration and Use Manual
MMI-20039472
Select the active concentration matrix using ProLink III
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
1. Choose Device Tools → Configuration → Process Measurement → Concentration Measurement.
The Concentration Measurement window is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Active Matrix to the matrix you want to use and select Change Matrix.
5.2.4 Set up concentration measurement using the Field
Communicator
This section guides you through most of the tasks related to setting up and implementing the concentration
measurement application.
Restriction
This section does not cover building a concentration matrix. See the Micro Motion Enhanced Density Application
Manual for detailed information on building a matrix.
Enable concentration measurement using the Field Communicator
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
The concentration measurement application cannot be enabled at the same time as the API referral
application or the Advanced Phase Measurement application. They must be disabled first.
Procedure
1. Choose Overview → Device Information → Licenses → Enable/Disable Applications and ensure that
Volume Flow Type is set to Liquid.
2. Choose Overview → Device Information → Licenses → Enable/Disable Applications.
3. Enable the concentration measurement application.
Load a concentration matrix from the transmitter's SD card using the Field
Communicator
If you have a concentration matrix on the transmitter's SD card, you can move it into one of the six slots on
your transmitter. You cannot use the matrix for measurement until it has been loaded into a slot. You can
load up to six matrices into slots.
Prerequisites
You must have one or more concentration matrices loaded onto the transmitter's SD card.
You must know the derived variable that the matrix is designed to calculate.
Configuration and Use Manual
Configure process measurement applications
June 2019
79