Adjust Concentration Measurement With Trim Slope And Trim Offset - Emerson Micro Motion 5700 Configuration And Use Manual

Transmitters with intrinsically safe outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configuration and Use Manual
MMI-20039472
You must be able to take a sample of your process fluid and obtain a laboratory concentration value at line
density and line temperature.
Procedure
1. Take a concentration reading from the meter, and record line density and line temperature.
2. Take a sample of the process fluid and obtain a laboratory value for concentration at line density and
line temperature, in the units used by the meter.
3. Subtract the meter value from the laboratory value.
4. Enter the result as the trim offset.
Using the display: Choose Menu → Configuration → Process Measurement → Concentration
Measurement → Configure Matrix, select your matrix, and enter Trim Offset.
Using ProLink III: Choose Device Tools → Configuration → Process Measurement →
Concentration Measurement, set Matrix Being Configured to your matrix, and enter Trim Offset.
Using the Field Communicator: Choose Configure → Manual Setup → Measurements → Optional
Setup → Conc Measurement → Configure Matrix and set Matrix Being Configured to your matrix.
Then choose Service Tools → Maintenance → Calibration → Trim CM Process Variables and enter
Concentration Offset.
5. Take another concentration reading from the meter, and compare it to the laboratory value.
If the two values are acceptably close, the trim is complete.
If the two values are not acceptably close, repeat this procedure.
Calculating the trim offset
Laboratory value
Meter value
Concentration offset: −0.72
13.8 Adjust concentration measurement with Trim Slope
and Trim Offset
Trim Slope and Trim Offset adjust the meter's concentration measurement to match a reference value.
Tip
You can adjust concentration measurement by applying the trim offset only, or by applying both the trim
offset and the trim slope. For most applications, the trim offset is sufficient.
Prerequisites
Ensure that the active matrix is the one that you want to trim. You can set the offset and slope separately for
each matrix on your transmitter.
You must be able to take measurements of your process fluid at two different concentrations.
You must be able to take a sample of your process fluid at each of these concentrations.
Configuration and Use Manual
64.21 °Brix
64.93 °Brix
64.21 − 64.93 = −0.72
Measurement support
June 2019
191