Troubleshooting
June 2019
16.10.50 mA Output 3 Fixed
Alert
Output Fixed
Cause
The output is configured to send a constant value. A loop test may be in progress.
Recommended actions
1. Check to see if a loop test is in process (the output is fixed).
2. Exit mA output trim, if applicable.
16.10.51 mA Output 3 Saturated
Alert
Output Saturated
Cause
The calculated output value is outside the range of the output.
Recommended actions
1. Check the settings of Upper Range Value and Lower Range Value.
Refer to
Configure Lower Range Value (LRV) and Upper Range Value (URV) for the mA
2. Check your process conditions against the values reported by the device.
3. Ensure that both devices are using the same measurement unit.
4. Purge the sensor tubes.
16.10.52 Moderate Phase Severity
Alert
Process Aberration
Cause
The transmitter has detected moderate two-phase severity.
Recommended actions
Check your process conditions against the values reported by the device.
16.10.53 Moderate Two Phase Flow
Alert
Process Aberration
244
Configuration and Use Manual
Micro Motion 5700 Transmitters with Intrinsically Safe Outputs
MMI-20039472
Output.