Configure device options and preferences
June 2019
2. If you turned on Auto Scroll, set Scroll Rate as desired.
Default: 10
Range: 1 to 30 seconds
Tip
Scroll Rate may not be available until you apply Auto Scroll.
7.1.5 Configure the display backlight
Display
Menu → Configuration → Display Settings
ProLink III
Device Tools → Configuration → Transmitter Display → General → Backlight
Field Communicator
Configure → Manual Setup → Display → Backlight
You can control the intensity and contrast of the backlight on the display's LCD panel.
Procedure
1. Set Intensity as desired.
Default: 50
Range: 0 to 100
2. Set Contrast as desired.
Default: 50
Range: 0 to 100
7.1.6 Configure security for the display
Menu → Configuration → Security → Display Security
Display
Device Tools → Configuration → Transmitter Display → Display Security
ProLink III
Configure → Manual Setup → Display → Display Menus
Field Communicator
When using the display, you can require users to enter a password to do any of the following tasks:
Enter the main menu
Change a parameter
Access alert data through the display
Start, stop, or reset totalizers or inventories via the context menu
The display password can be the same or different from the totalizer/inventory context menu control
password. If different, the display password is used to reset, start, and stop totalizers or inventories using
Menu → Operations → Totalizers.
Procedure
1. Configure Password Required as desired.
114
Configuration and Use Manual
Micro Motion 5700 Transmitters with Intrinsically Safe Outputs
MMI-20039472

Advertisement

Table of Contents
loading