Emerson Micro Motion 5700 Configuration And Use Manual page 178

Transmitters with intrinsically safe outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Measurement support
June 2019
Zero Time controls the amount of time the transmitter takes to determine its zero-flow reference
point. The default Zero Time is 20 seconds. For most applications, the default Zero Time is
appropriate.
3. Start the zero procedure and wait until it completes.
When the calibration is complete:
If the zero procedure was successful, a Calibration Success message and a new zero value are
displayed.
If the zero procedure failed, a Calibration Failed message is displayed.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Need help?
If the zero fails:
Ensure that there is no flow through the sensor, then retry.
Remove or reduce sources of electromechanical noise, then retry.
Set Zero Time to a lower value, then retry.
If the zero continues to fail, contact customer service.
If you want to restore the most recent valid value from transmitter memory:
Using the display: Menu → Service Tools → Verification and Calibration → Meter Zero → Restore
Zero → Restore Previous Zero
Using ProLink III: Device Tools → Calibration → Smart Zero Verification and Calibration → Calibrate
Zero → Restore Prior Zero
Using the Field Communicator: Not available
If you want to restore the factory zero:
Using the display: Menu → Service Tools → Verification and Calibration → Meter Zero → Restore
Zero → Restore Factory Zero
Using ProLink III: Device Tools → Calibration → Smart Zero Verification and Calibration → Calibrate
Zero → Restore Factory Zero
Using the Field Communicator: Service Tools → Maintenance → Calibration → Zero Calibration →
Restore Factory Zero
Restriction
Restore the factory zero only if your meter was purchased as a unit, it was zeroed at the factory, and you are
using the original components.
Related information
Verify the zero
178
Configuration and Use Manual
Micro Motion 5700 Transmitters with Intrinsically Safe Outputs
MMI-20039472

Advertisement

Table of Contents
loading

Table of Contents