Density Measurement Problems - Emerson Micro Motion 5700 Configuration And Use Manual

Transmitters with intrinsically safe outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Troubleshooting
June 2019
16.4.2 Inaccurate CM reading
Cause
Inaccurate density measurement
Inaccurate temperature measurement
Incorrect reference conditions
Incorrect matrix data
Inappropriate trim values
Recommended actions
1. Verify the line density value.
2. Verify the line temperature value.
3. Ensure that the application is configured to use the appropriate temperature source.
4. Ensure that reference temperature is configured correctly.
5. Ensure that the appropriate matrix is active.
6. Ensure that the matrix is configured correctly.
7. Adjust the extrapolation limits for the active matrix.
8. Adjust measurement with a concentration offset trim.

16.5 Density measurement problems

16.5.1 Erratic density reading
Cause
Normal process noise
Two-phase flow
Line pressure too low
The flow rate is too high for the installation
Pipe diameter too small
Contaminants or suspended solids in the process gas
Contaminants or suspended solids in the process fluid
Vibration in the pipeline
Erosion or corrosion
Recommended actions
1. Check your process conditions against the values reported by the device.
2. Increase the density damping value.
3. Decrease the flow rate.
4. Check for two-phase flow.
216
Configuration and Use Manual
Micro Motion 5700 Transmitters with Intrinsically Safe Outputs
MMI-20039472