Troubleshooting
June 2019
16.10.3 [005] Mass Flow Rate Overrange
Alert
Extreme Primary Purpose Variable
Cause
The measured flow is outside the sensor's flow limits.
Recommended actions
1. Check your process conditions against the values reported by the device.
2. Verify that the transmitter is configured correctly for the connected sensor.
3. Check for two-phase flow.
Refer to
Check for two-phase flow (slug
4. Contact customer service.
16.10.4 [008] Density Overrange
Alert
Extreme Primary Purpose Variable
Cause
The measured density is above 10 g/cm
Recommended actions
1. Check your process conditions against the values reported by the device.
2. Verify that the transmitter is configured correctly for the connected sensor.
3. Check for two-phase flow.
4. Verify all of the characterization or calibration parameters.
See the sensor tag or the calibration sheet for your meter.
5. Check the sensor coils for electrical shorts. If you find problems, replace the sensor.
6. Check the drive gain and the pickoff voltage.
7. Contact customer service.
16.10.5 [009] Transmitter Initializing/ Warming Up
Alert
Transmitter Initializing
Cause
The transmitter is in power-up mode.
Recommended actions
1. Allow the meter to complete its power-up sequence. The alert should clear automatically.
228
flow).
3
.
Micro Motion 5700 Transmitters with Intrinsically Safe Outputs