Understanding the Multidrop Pico dispenser
Adjusting the
It is important to set the height of the cassette above the well plate so
that it clears the well plate and dispenses optimally.
cassette-to-plate
Turn off the Multidrop Pico dispenser.
height
Place a well plate on the destination plate holder and move it under the
cassette holder.
Use a cassette as a spacer by placing it directly on the well plate. The
cassette should be placed as shown below.
Loosen the adjustment knob and lower the cassette holder until it
touches the cassette. The space should be ~1.25 mm. Gently tighten the
knob and remove the cassette.
The height adjustment can also be made at protocol runtime before the
actual dispensing or during A1 alignment check before actual
dispensing.
17