General Operational Instructions

Switching the Instrument On

1. To switch on the instrument, press the 'On/Off ' key briefly. There will be a short beep and a slight
pause followed by the Geotech logo.
2. The power on self-test will then commence (approximately 15 seconds) including warm-up time.
3. Assuming there are no warnings to display the instrument will continue to the 'Main Read Screen'.

Switching the Instrument Off

1. Purge with fresh air. Run the pump for approximately 30 seconds or until the readings have
returned to normal levels.
2. To switch off the instrument, press the 'On/Off ' key briefly.

Instrument Main Read Screen

After the analyser has been switched on and the warm-up self-tests completed the analyser will display
the 'Main Read Screen'.
Soft-Keys:
Peak - Enables the operator to display the peak reading.
Store - Enables the operator to store the reading for viewing/download later.
Hold - Enables the operator to hold the current reading being taken.
Thermo Scientific
Note Before the instrument is switched off a clean air purge should be performed. This
ensures that the instrument is free from gas and ready for the next measurement. This final
purge is especially important for the oxygen sensor as it may degrade if stored when
contaminated with gas.
Note If the 'Auto Off ' utilities setting is set to 'Yes', the analyser will switch off
automatically after 10 minutes if not in use.