Floating - Keysight Technologies B2961A User Manual

Low noise power source
Hide thumbs Also See for B2961A:
Table of Contents

Advertisement

Figure 3-3
Installation
Connecting a DUT
Dual View, 4-Wire Connection Setup for Channel 1

Floating

With the default setting, Low Force terminal is connected to the chassis ground.
However, it can be internally disconnected from the ground for making the floating
condition. This setup is used for connecting the source channels in series.
To specify the low terminal status, perform the following procedure. Setting is not
saved. Turning off the instrument sets the status to the grounded.
1. Press the Config > Source > Connection function keys. This opens the Output
Connection dialog box.
2. Set the Low Terminal State on this dialog box. It must be FLOATING for the
floating status, or GROUNDED for the grounded status.
3. Press the OK softkey.
For the floating status, the following indicator appears on the status information
area. And then the Low Force and Low Sense terminals can be connected to the
maximum of 250 V.
Channel 1 floating status indicator:
Channel 2 floating status indicator:
3-16
Keysight B2961A/B2962A User's Guide, Edition 3