Keysight Technologies B2961A Demo Manual

Keysight Technologies B2961A Demo Manual

Sourcing precise current/voltage sinusoidal waveforms using 6.5 digit low noise power source
Hide thumbs Also See for B2961A:

Advertisement

Quick Links

Keysight Technologies
Sourcing Precise Current/Voltage Sinusoidal
Waveforms Using the B2961A/62A
B2961A/62A 6.5 Digit Low Noise Power Source
Demo Guide

Advertisement

Table of Contents
loading

Summary of Contents for Keysight Technologies B2961A

  • Page 1 Keysight Technologies Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A B2961A/62A 6.5 Digit Low Noise Power Source Demo Guide...
  • Page 2: Table Of Contents

    B2961A/62A Setup ........
  • Page 3: Introduction

    All of the accessories required to perform the primary demos described in this demonstration guide are provided in a demo kit that is included with Keysight B2961A/62A demo units. The kit includes items such as a banana cable, a 1.1 kΩ resistor, etc., so the only additional item needed is an oscilloscope.
  • Page 4: Concept

    04 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide Concept Figure 1 illustrates the connection diagram and basic setup conditions used in the demo to source current/voltage sinusoidal waveforms using the B2961A/62A. The B2961A/62A will source a sinusoidal waveform using the settings shown below while the output is monitored using an oscilloscope.
  • Page 5: Setup

    05 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide Setup Oscilloscope Setup 1. Set Ch1 vertical scale to 500 mV/div and horizontal scale to 5 ms/div. 2. Set Ch1 Input Impedance to 1M Ohm.
  • Page 6: Lab1: Source Finite Current/Voltage Sinusoid Waveform

    06 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide LAB1: Source Finite Current/Voltage Sinusoid Waveform Demonstration Objective 1-1. Source a finite current/voltage sinusoid waveform This demo illustrates the arbitrary 1) Press repeatedly until the Channel 1 Single View is displayed.
  • Page 7 07 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide 6) Press to toggle the Assist keys, and press to display the Waveform Preview. This enables you to verify the waveform configuration on the instrument’s GUI.
  • Page 8 To confirm the output voltage waveform, perform the following procedure. 13) Rotate and press to edit Y-axis data type, and select to set the Y-axis data type to Voltage. Now you can see the monitored output voltage waveform on the B2961A/62A’s GUI as shown below Result on B2961A/62A’s GUI...
  • Page 9 09 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide Result on Oscilloscope Why use the B2960A Series’ Arbitrary Waveform Generation Function? Some conventional power supplies can also generate waveforms but the signal integrity is not good.
  • Page 10: Optional) Configure Phase Marker Output

    10 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide 1-2. (Optional) Configure phase marker output If desired you can enable the phase marker output function to define a specific point on a waveform anywhere from 0 to 360 degrees. The phase marker output signal is sourced through the digital I/O port.
  • Page 11 11 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide 5) Press , and then press to open the DIO Configuration dialogue. (1) Press More... (2) Press I/O (3) Press DIO (4) Press Config 6) Press and select to specify the Digital I/O pin number;...
  • Page 12: Monitor The Entire Output Signal On Its Gui

    Oscilloscope waveforms 1-3. Monitor the entire output signal on its GUI The default settings of the B2961A/62A make it monitor only the first waveform even if multiple waveforms are configured on the instrument. However, you can modify the Trigger Parameters to monitor the B2961A/62A‘s entire output and eliminate the need for an oscilloscope.
  • Page 13 13 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide 3) Press to edit the Trigger type, and then select to set the Trigger type to TIMER; this will make a time domain measurement. 4) Rotate to select the Trigger Parameters for Channel 1 and configure them as shown below.
  • Page 14: View The List Of Measurement Results

    14 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide 1-4. View a list of the measurement results The measurement results including the measurement time stamp can be referred by the following steps. 1) If you are not on the top of the Function menu, press repeatedly until you are back at the top level.
  • Page 15: Lab2: Source Continuous Current/Voltage Sinusoid Waveform

    15 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide LAB2: Source a continuous current/voltage sinusoid waveform Demonstration Objective 2-1. Source continuous current/voltage sinusoid waveform To source a continuous waveform, perform the following procedure. This demo shows how to source a continuous (infinite) waveform.
  • Page 16: Stop Sourcing Continuous Current/Voltage Sinusoid Waveform

    16 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide 2-2. Stop sourcing continuous current/voltage sinusoid waveform If you are sourcing a continuous waveform and you want to stop sourcing it, perform the following procedure.
  • Page 17: Lab3: Sourcing Current/Voltage Sinusoid Waveform Without Monitoring

    17 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide LAB3: Sourcing current/voltage sinusoid waveform without monitoring Demonstration Objective To source a waveform without monitoring it, perform the following procedure to execute only a Transient (Source) Action.
  • Page 18: Lab4: (Optional) Source Current/Voltage Sinusoid Waveform When The B2961A/62A Receives An External Trigger Signal

    18 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide LAB4: (Optional) Source current/voltage sinusoid waveform when the B2961A/62A receives an external trigger signal Demonstration Objective To set up the instrument to source a current/voltage sinusoid waveform when the...
  • Page 19 19 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide 2) Press to show the Trigger Sub-Panel. 3) Press to edit Trigger type, and then select to set Trigger type to MANUAL. 4) Rotate to select the Source trigger source and press to edit it.
  • Page 20 20 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide 7) Press , and then press to open the DIO Configuration dialogue. (1) Press More... (2) Press I/O (3) Press DIO (4) Press Config 8) Press and select to specify the Digital I/O pin number;...
  • Page 21 21 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide 12) Rotate to select Source trigger count and press to edit it. Then enter the number of trigger signals that will be sent to the instrument as Source trigger count.
  • Page 22 22 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide To return the instrument to an idle state before completing all waveform sourcing, perform the following procedure. 15) If you are not on the top Function menu, press repeatedly until you are back at the top level.
  • Page 23 23 | Keysight | Sourcing Precise Current/Voltage Sinusoidal Waveforms Using the B2961A/62A – Demo Guide From Hewlett-Packard through Agilent to Keysight For more information on Keysight For more than 75 years, we‘ve been helping you unlock measurement insights. Technologies’ products, applications or...