6.1.2. L1 PHASE VOLTAGE HARMONICS DISPLAY SCREEN
Information about the harmonic at
the cursor position.
V-h03: harmonic number.
%: level of harmonics with the
fundamental RMS value as
reference (%f) or the (total) RMS
value as reference (%r).
V: RMS voltage of the harmonic
in question.
-143°: phase shift with respect to
the fundamental (order 1).
max – min: maximum and
minimum levels of the harmonic
in question. They are reset
when the harmonic number is
changed or the button is
THD: total harmonic distortion.
Vd: RMS distortion voltage.
NOTES: Filters L2 and L3 display the harmonics of the phase-to-neutral voltage for phases 2 and 3, respectively. The
screen is identical to the one displayed for filter L1.
There is no phase shift or distortion value for the neutral channel.
6.2. CURRENT
The A sub-menu displays harmonics of the current.
6.2.1. 3L CURRENT HARMONICS DISPLAY SCREEN
Information about the harmonic
at the cursor position.
A-h05: harmonic number.
%: level of harmonics with the
fundamental RMS value as
reference (%f) or the (total) RMS
value as reference (%r).
A: RMS current of the harmonic
in question.
+179°: phase shift with respect to
the fundamental (order 1).
Harmonic selection cursor. Use the
or button to move the cursor.
Harmonic selection
pressed.
cursor. Use the or
button to move the cursor
Indicator of the
presence of non-zero
harmonics of order
higher than 25
Figure 6-3
Figure 6-4
49
Display of the 3 phases 3L, of L1, L2,
L3, N, or of the expert mode (three-
phase connection only - see §6.5). To
select the display press or .
The horizontal axis indicates the
orders of the harmonics. The level of the
harmonic is given as a percentage with
respect to the fundamental or to the
total RMS value.
DC: DC component.
1 to 25: harmonics of order 1 to 25.
When the cursor exceeds order 25,
order 26 to 50 appears.
Display of the 3 phases 3L, of L1, L2,
L3, N, or of the expert mode (three-
phase connection only – see §6.5). To
select the display press or .
The horizontal axis indicates the orders
of the harmonics. Harmonic level is
shown as a percentage with respect to
the fundamental or to the total RMS.
Rang DC: DC component.
1 to 25: harmonics of order 1 to 25.
When the cursor exceeds order 25,
order 26 to 50 appears.