3.1. START-UP

To turn the instrument ON, press the
After the software check, the home page is displayed, followed by the information screen that indicates the software version of the
instrument and its serial number. The Waveform screen is then displayed:

3.2. CONFIGURATION

To configure the instrument:
1.
Press
. The Configuration screen appears (see Figure 4-1).
2.
Press  or  to select the parameter to be modified.
3.
Press  to enter the selected sub-menu.
4.
Use the arrow buttons ( or  and  or ) and the  button to validate. For more details, see §§4.3 to 4.10.
The following settings must be checked (and if necessary modified) for each measurement:
Calculation methods (see §4.5).
Distribution system (single-phase to five-wire three-phase) and connection method (2 wattmeters, 2 ½ elements, standard)
(§4.6).
Current ratios according to the type of current sensor connected (§4.7).
Voltage ratios (§4.7).
Triggering levels (transients mode and inrush current capture) (§4.8).
Values to be recorded (trend mode) (§4.9).
Alarm thresholds (§4.10).
To return to the Configuration screen from a sub-menu, press the
3.

OPERATION

button.