Instantaneous Inrush Current - AEMC PowerPad 8436 User Manual

3-phase power quality analyzer
Hide thumbs Also See for PowerPad 8436:
Table of Contents

Advertisement

5.2.4. INSTANTANEOUS INRUSH CURRENT

PEAK mode displays the envelopes and waveforms of the inrush current capture. The PEAK display of an inrush current capture
provides two possible representations:
envelope
waveform
The change from one of these representations to the other is automatic, as a function of the zoom level. If the zoom in is high
enough, the representation is of the "waveform" type.
The display depends on the type of display filter:
4V: displays the 4 voltages during the inrush current capture for set-ups with neutral (for a waveform type display only).
3U: displays the 3 voltages during the inrush current capture for set-ups without neutral (for a waveform type display only).
4A: displays the 4 currents during the inrush current capture (for a waveform type display only).
L1, L2 or L3: display the voltage and current of phases 1, 2, and 3, respectively (only for set-ups with neutral and for a waveform
type display).
N: displays the neutral current and neutral voltage during the inrush current capture (for a waveform type display only).
V1, V2, V3: displays the 3 voltages during the inrush current capture for set-ups with neutral (for an envelope type display only).
U1, U2, U3: displays the 3 voltages during the inrush current capture for set-ups without neutral (for an envelope type display only).
A1, A2, A3: displays the 3 currents during the inrush current capture (for an envelope type display only).
Below are three examples of display.
5.2.4.1. 4A PEAK display screen for a 5-wire three-phase connection
Location of the zone displayed in
MAX |PEAK|: maximum
instantaneous absolute value of
the inrush current capture.
Scale of values in amperes.
t: relative time position of the cursor
(t=0 corresponds to the start of the
inrush current capture).
5.2.4.2. 3A PEAK display screen for a 3-wire three-phase connection
the record.
Figure 5-12
Figure 5-13
46
Number assigned to the displayed
graph. In this example, circle 3 is filled
in to indicate channel A3 triggered
capture of the inrush current.
Time cursor. Use the  or  button to
move the cursor
A1, A2, A3: instantaneous values of
currents 1, 2, and 3 at the position of
the cursor

Advertisement

Table of Contents
loading

Table of Contents