Cleaning Requirements - Welch Allyn Connex Spot Monitor Service Manual

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

Cleaning requirements

This section presents procedures for cleaning the Connex Spot Monitor (including the
monitor, stand, APM work surface, accessories, and accessory basket and bins).
Welch Allyn has validated these instructions to be capable of preparing your Connex
Spot Monitor devices and above accessories for re-use. Clean on a routine basis
according to your facility's protocols and standards or local regulations. If the monitor is
on, lock the display.
If liquids are spilled on the monitor:
1. Power down the monitor.
2. Disconnect the power cord from the mains outlet and the power source.
3. Remove battery pack from the monitor.
4. Dry off excess liquid from the monitor.
Note
5. Reinstall battery pack.
6. Reconnect the power cord.
7. Power on the monitor and verify that the monitor functions normally before using it.
WARNING Patient injury risk. Clean all accessories, including cables and
tubes, before storing the accessories on the device or stand. This helps
reduce the risk of cross contamination and nosocomial infection.
WARNING Electric shock hazard. Before cleaning the monitor, disconnect
the AC power cord from the mains outlet and the power source.
WARNING Electric shock hazard. DO NOT immerse or autoclave the
monitor or accessories. The monitor and the accessories are not heat-
resistant.
WARNING Liquids can damage electronics inside the monitor. Prevent
liquids from spilling on the monitor.
CAUTION Do not sterilize the monitor. Sterilizing the monitor could harm
the device.
If liquids possibly entered the monitor, remove the monitor
from use until it has been properly dried, inspected, and
tested by qualified service personnel.
93