200 Appendix
APM messages
Message
APM not functional.
1C1001
APM not functional.
1C100B
APM battery is absent
or faulty. 1C100F
The APM is
disconnected. 1C1002
USB cable is
disconnected. 1C1003
APM is plugged in.
APM not functional.
1C1010
APM not functional.
1C1004
APM not functional.
1C1009
Possible cause
The APM is detected as connected
but there is no communication
through the APM serial port
The APM battery is installed, but
does not communicate with the
monitor
The APM battery is not installed
The APM is unplugged from the
monitor while the monitor is
powered on
The APM USB hub is unplugged
from the monitor while the
monitor is powered on
The APM was plugged in while
the monitor is powered on.
The APM USB hub is plugged in
while the monitor communication
cable is disconnected
The APM PIC cannot communicate
with the accelerometer
The APM PIC software update and
any retries have failed
Suggested action
Restart the monitor and the APM. If the
problem is still present, check the cable
connections from the monitor to APM. If the
problem still present, replace the APM. If the
message persists, replace the main PCBA on
the monitor.
Perform diagnostic checks on the monitor. If
the problem is still present, replace the APM
battery. If the problem still present, replace
the APM. If the message persists, replace
the main PCBA on the monitor.
Ensure than an APM battery is installed, and
install one if it is missing. If the problem is
still present, perform diagnostic checks on
the monitor. If the problem is still present,
replace the APM. If the message persists,
replace the main PCBA on the monitor.
Restart the monitor and the APM. If the
problem is still present, check the cable
connections from the monitor to APM. If the
problem still present, replace the APM. If the
message persists, replace the main PCBA on
the monitor.
Restart the monitor and the APM. If the
problem is still present, check the cable
connections from the monitor to APM. If the
problem still present, replace the APM. If the
message persists, replace the main PCBA on
the monitor.
Restart the monitor and the APM. If the
problem is still present, check the cable
connections from the monitor to APM. If the
problem still present, replace the APM. If the
message persists, replace the main PCBA on
the monitor.
Restart the monitor and the APM. If the
problem is still present, check the cable
connections from the monitor to APM. If the
problem still present, replace the APM. If the
message persists, replace the main PCBA on
the monitor.
Restart the monitor and the APM. If the
problem is still present, check the cable
connections from the monitor to APM. If the
problem still present, replace APM. If the
message persists, replace the main PCBA on
the monitor.
Restart the monitor and the APM. If the
problem is still present, check the cable
®
Connex
Spot Monitor
Alarm
priority
Very low
Very low
Very low
Very low
Very low
Information
Very low
Very low
Very low