124 Functional verification and calibration overview
Material no.
N/A
06138-000
N/A
DOC-10
106270
N/A
106275
6000-30
N/A
N/A
1001120
web download
4500-30
620216
Initialize the NIBP board
Functional verification and calibration must be performed each time the device is
disassembled or more frequently if your facilty's procedures require a full calibration on a
specified basis. If you have just replaced a Main PCA board, your first step before
performing a full verification and calibration is to initialize the NIBP Board.
1. Set up the equipment for full functional verification as shown earlier in this section.
2. Log in to the Welch Allyn Service Tool, highlight your device in the Device list, and
click Select.
This action opens the Connex Spot Monitor > Device information subtab.
Depending on your configuration, it could take 30-45 seconds to populate this
subtab.
3. Select the Verify and calibrate subtab.
4. Select NIBP sensor in the Device field.
5. Click Calibrate.
Description
SpO2 simulator
Cal-key, assembly, M690/692
Pressure meter (must include at least two decimal points and be accurate to
within ±0.5 mmHg)
Nellcor SpO2 extension cable
Service test box
Ethernet cable
Micro USB cable
Single tube blood pressure hose, 5 feet
Wireless router, a/b/g or equivalent
PC running Windows XP SP3, Windows 7, or Windows 8
Welch Allyn Service Tool Gold license
Welch Allyn Service Tool
welchallyn.com/en/service-support/service-center/service-tool.html
Blood pressure hose, 5 ft.
Fitting "Y" 1/8 X 1/8 X 1/8
®
Connex
Spot Monitor
Qty Component
1
SpO2
1
SureTemp Plus
Thermometry
Module
1
NIBP
1
Nellcor SpO2
1
Nurse Call, USB
ports
1
Ethernet Test
1
All
1
NIBP
1
Wired &
Wireless Test
1
All
1
NIBP, Software
updates
1
NIBP, Software
updates and
calibration
1
NIBP
1
NIBP