Message
Radio not functional.
35000f
Radio not functional.
350010
Radio not functional.
350011
Invalid radio
configuration.
Reconfigure and try
again. 350012
Invalid radio
configuration.
Reconfigure and try
again. 350013
Invalid radio
configuration.
Reconfigure and try
again. 350014
Radio not functional.
350015
Radio not functional.
350016
Radio not functional.
350017
Invalid radio
configuration.
Reconfigure and try
again. 350018
Invalid radio
configuration.
Reconfigure and try
again. 350019
Possible cause
No statistics file. There is an
internal software error on the
radio indicating a Linux kernel
error
Missing interface. There is an
internal software error on the
radio indicating a Linux kernel
error or failure to initialize the
network interface
Unknown interface. There is a
software communication issue
between the host and the radio
Invalid WEP key length. WEP keys
should be 10 or 26 characters
long.
Not in EAP mode. There is an
internal software error on the
monitor: attempt to configure
settings that don't apply in current
authentication mode on radio
Invalid inner EAP method. There is
an internal software error on the
monitor: attempt to configure
settings that don't apply in current
authentication mode on radio
Out of memory. There is an
internal software error on the
radio
Invalid log level. There is a
software communication issue on
the radio
Certificate path too long. There is
an internal software error on the
radio. The radio has a fixed path
length
Missing client certificate. The
radio attempted to configure for
an EAP mode that requires a client
certificate and no certificate is
installed
Missing CA certification. The radio
attempted to enable server
validation and CA certification is
missing
Suggested action
Check for a software update and apply it.
If still present, replace the radio.
Check for a software update and apply it.
If still present, replace the radio.
Check for a software update and apply it.
If still present, replace the radio.
Check the radio configuration. If the
problem is still present, reset the radio to
factory defaults. If the problem persists,
check for software update and apply it. If
still present, replace the radio.
Check the radio configuration. If the
problem is still present, reset the radio to
factory defaults. If the problem persists,
check for software update and apply it. If
still present, replace the radio.
Check the radio configuration. If the
problem is still present, reset the radio to
factory defaults. If the problem persists,
check for software update and apply it. If
still present, replace the radio.
Check for a software update and apply it.
If still present, replace the radio.
Check for a software update and apply it.
If still present, replace the radio.
Check for a software update and apply it.
If still present, replace the radio.
Check the radio configuration. If the
problem is still present, reset the radio to
factory defaults. If the problem persists,
check for software update and apply it. If
still present, replace the radio.
Check the radio configuration. If the
problem is still present, reset the radio to
factory defaults. If the problem persists,
check for software update and apply it. If
still present, replace the radio.
®