Welch Allyn Connex Spot Monitor Service Manual page 187

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

Service manual
Message
Temperature not functional.
3080A
Temperature not functional.
3080B
Replace temperature probe.
3080C
Temperature not functional.
3080D
Temperature not functional.
3080E
Connect temperature probe.
3080F
Replace temperature probe.
30810
Temperature not functional.
30811
Temperature not functional.
30812
Replace temperature probe. Code
30813
Temperature not functional.
30814
Possible cause
Suggested action
The temperature
Internal malfunction. If the problem
module has a problem
persists, replace the module.
saving to the monitor
EEPROM in biotech
mode
The temperature
Internal malfunction. If the problem
module error detection
persists, replace the module.
mechanism detected an
error
The temperature
Probe malfunction. Replace probe. If the
module probe error
problem persists, replace the module.
detection mechanism
detected an error
The temperature
Probe malfunction. Replace probe. If the
module log error
problem persists, replace the module.
detection mechanism
detected an error
The temperature
Probe malfunction. Replace probe. If the
module calibration error
problem persists, replace the module.
detection mechanism
detected an error
The temperature
Probe malfunction. Replace probe. If the
module detected no
problem persists, replace the module.
probe connected
The temperature
Probe malfunction. Replace probe. If the
module cannot read the
problem persists, replace the module.
probe EEPROM
correctly or the probe
left the factory without
being tested.
The temperature
Internal malfunction. If the problem
module has an invalid
persists, replace the module.
events index
There is a problem
Internal malfunction. If the problem
reading the
persists, replace the module.
temperature module
EEPROM or saving to
the monitor EEPROM in
biotech mode.
The temperature
Probe malfunction. Replace probe. If the
module has a problem
problem persists, replace the module.
reading the probe
EEPROM.
The temperature
Internal malfunction. If the problem
module TEMP CONFIG
persists, replace the module.
ACQUIRE FAILURE
Appendix 181
Alarm priority
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low