Welch Allyn Connex Spot Monitor Service Manual page 205

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

Service manual
®
Bluetooth
messages
Message
Bluetooth connection
failure. 370201
Bluetooth not functional.
370001
Bluetooth not functional.
370002
N/A
N/A
Bluetooth device
connection successful
N/A
Bluetooth device
disconnected
Possible cause
The monitor failed to connect to
the Bluetooth device.
The monitor detected a Bluetooth
device that is not functional
The monitor cannot detect a
Bluetooth module
The Bluetooth device failed to pair
with the monitor.
The Bluetooth device failed to pair
with the monitor.
The Bluetooth device connected
The Bluetooth device did not
connect with the monitor
The Bluetooth disconnected
Suggested action
Verify a successful pairing of the device
with the PC. If the problem is still present,
verify the EMR and the patient window are
running. If the problem is still present,
verify the Bluetooth device signal strength.
If the problem is still present, verify the PC
hardware and configuration. If the problem
is still present, verify the Bluetooth device
hardware.
Reboot the monitor. If the problem persists,
replace the Bluetooth radio. If the problem
persists, replace the main PCBA.
Replace the Bluetooth radio. If the problem
persists, replace the main PCBA.
Verify a successful pairing between the
device and the PC. If the problem is still
present, verify the Bluetooth signal
strength. If the problem persists, verify
authentication code, verify the computer
hardware and configuration. If the problem
persists, verify if the computer has prior
pairing information for the monitor.
Version of stack on the PC does not support
simple secure pairing. The device requires
2.1+EDR Bluetooth stack on the host PC.
None.
BT device connection dropped due to
distance, application crash on host end.
OR
RF connection not possible
OR
RF connection possible, but cannot find
device
OR
BT connection was established, but WACP
connection not possible
None.
Appendix 199
Alarm
priority
Very low
Very low
Very low
Very low
Very low
Information
N/A
Information

Hide quick links: