Identifying The Monitor And Subsystems - Welch Allyn Connex Spot Monitor Service Manual

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

Service manual
Message
APM not functional.
1C100B
APM not functional.
Device is operating in
battery mode.
Sleep mode is
unavailable. Intervals
monitoring is in
progress.
Sleep mode is
unavailable. An alarm is
active.
Display lock is
unavailable. Missing
patient context.
Power cable is
disconnected. 1C1011

Identifying the monitor and subsystems

This explanation of the monitor labeling is provided to assist you in identifying these
particular parts by serial number. The part numbers appearing on the labels are internal
part numbers and will not be found in the list of field replaceable units (FRUs) included in
this manual.
Future revisions of a part may not always be backward compatible with a legacy part. To
determine the correct replacement part for your unit always provide the serial number
and model when ordering.
Possible cause
The APM battery is not recharging Restart the monitor and the APM. If the
The APM USB changes from
unplugged to plugged after
monitor startup
AC power cord has been
disconnected.
Sleep mode is not allowed when
intervals are in progress
Sleep mode is not allowed when
alarms are active
Lockout is not allowed without
active patient information
The APM communication cable is
plugged in while the APM USB
cable is disconnected
Suggested action
connections from the monitor to APM. If the
problem still present, retry the software
update. If still present, replace the APM. If
the message persists, replace the main PCBA
on the monitor.
problem is still present, check the cable
connections from the monitor to APM. If the
problem still present, replace the APM. If the
message persists, replace the main PCBA on
the monitor.
Restart the monitor and the APM. If the
problem is still present, check the cable
connections from the monitor to APM. If the
problem still present, retry the software
update. If still present, replace the APM. If
the message persists, replace the main PCBA
on the monitor.
Information status message; press OK button
to dismiss.
Stop any active intervals.
Clear all active alarms.
Enter patient information
Restart the monitor and the APM. If the
problem is still present, check the cable
connections from the monitor to APM. If the
problem still present, retry the software
update. If still present, replace the APM. If
the message persists, replace the main PCBA
on the monitor.
Appendix 201
Alarm
priority
Very low
Information
Information
Information
Information
Information
Very low