Service manual
Message
Ambient temperature out of
range. Clear to retry. 30822
Temperature not functional.
30823
Temperature not functional.
30824
Temperature not functional.
30825
Temperature not functional.
30826
Temperature not functional.
30827
Temperature not functional.
30828
Temperature not functional.
30829
Temperature not functional.
3082A
Temperature not functional.
3082B
Temperature not functional.
03C800
Temperature not functional.
03C900
Temperature not functional.
03CA00
Possible cause
Suggested action
problem persists, replace the probe. If the
problem still persists, replace the module.
The temperature
Verify conditions are greater than 50 F or
module under ambient
10°C. If conditions are valid and the
temp
problem persists, replace the probe. If the
problem still persists, replace the module.
The temperature
Internal malfunction. If the problem
module has an invalid
persists, replace the module.
SureTemp algorithm
The temperature
Internal malfunction. If the problem
module is over max
persists, replace the module.
battery volts
The temperature
Internal malfunction. If the problem
module is under min
persists, replace the module.
battery volts
The temperature
Internal malfunction. If the problem
module battery volts
persists, replace the module.
not set
The temperature
Internal malfunction. If the problem
module predict
persists, replace the module.
algorithm is not set
The temperature
Internal malfunction. If the problem
module ambient temp
persists, replace the module.
is not set
The temperature
Probe malfunction. Replace probe. If the
module has a non-
problem persists, replace the module.
responsive probe. The
thermistor has pulled
away from the tip or
the heater is broken.
The temperature
Probe malfunction. Replace probe. If the
module is experiencing
problem persists, replace the module.
bad probe gain
The temperature
Probe malfunction. Replace probe. If the
module has a bad probe
problem persists, replace the module.
response value
The temperature
Internal malfunction. If the problem
module is not
persists, replace the module.
functional
Unable to deserialize
Internal malfunction. If the problem
messages from the
persists, replace the module.
temperature module
Unsupported message
Internal malfunction. If the problem
received from the
persists, replace the module.
Temperature module
Appendix 183
Alarm priority
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low