164 Appendix
Message
NIBP not calibrated.
Calibrate the module.
050509
Invalid algorithm. Select
correct algorithm and
retry. 05050A
NIBP not functional.
050513
Invalid patient mode.
Select correct patient
mode and retry. 050514
NIBP not functional.
050515
NIBP not functional.
050516
Ambient temperature
out of range. Clear error
and retry. 050517
Low battery. Plug into
outlet. 050518
Battery overcharged.
Disconnect from outlet.
050519
NIBP not calibrated.
Calibrate the module.
050601
NIBP not functional.
050602
NIBP not calibrated.
Calibrate the module.
050603
Cuff pressure limits
exceeded. 050604
Premature auto cycle
skipped. 050605
Cuff pressure too high.
Clear error to retry.
050606
Possible cause
NIBP module calibration failure,
the calibration signature is zero
Invalid NIBP Algorithm. NIBP
component software tried to
configure the sensor in an illegal
manner
Invalid NIBP initiation code
Invalid patient mode on NIBP. The
NIBP component software tried to
configure the sensor in an illegal
manner
Invalid module configuration for
NIBP
NIBP module malfunction
Ambient temp out of range on
NIBP
The NIBP module power rail is too
low
The NIBP module power rail is too
high. NIBP was calibrated without
A/C connected or battery too low.
NIBP failed to load safety
processors calibration record from
EEPROM
NIBP safety processor failed ROM
checksum
NIBP safety processor not
calibrated, missing calibration
signature
NIBP system failure. Overpressure Restrict patient movement.
NIBP auto cycle skipped, SVRP
requirement not satisfied
NIBP cuff pressure above SVRP for
to long
Suggested action
Calibrate the NIBP module.
Verify the algorithm. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Verify correct patient mode. If the
problem persists, replace the
module.
Internal malfunction. If the problem
persists, replace the module.
Internal malfunction. If the problem
persists, replace the module.
Restore device to normal
temperature ranges and try again.
Plug device into an AC outlet to
charge the battery.
Battery is overcharged. Remove from
charging source. Charge battery and
initialize NIBP module and
recalibrate with A/C connected.
Calibrate the NIBP module. If the
problem persists, replace the
module.
Internal malfunction. If the problem
persists, replace the module.
Calibrate the NIBP module. If the
problem persists, replace the
module.
Cuff pressure is not below safety
return pressure long enough to allow
a cycle to occur.
Verify cuff connections. If the
problem persists, replace the
module.
®
Connex
Spot Monitor
Alarm priority
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Very low
Medium
Very low
Medium