Welch Allyn Connex Spot Monitor Service Manual page 193

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

Service manual
Message
Unable to identify
clinician. Security
provider error.
Unable to identify
clinician. User not
found.
Unable to identify
clinician. Invalid ID or
system password.
Unable to identify
clinician. Account
disabled/expired.
Unable to identify
clinician. Password
expired/reset required.
Unable to identify
clinician. Group
membership error.
Unable to identify
clinician.
Touch Clear to delete all
data.
Unable to identify
patient.
Touch Clear to delete all
data.
Database schema out of
data; recreating.
Database is unreadable
during startup;
recreating. 1F0001
Error accessing PDM
database; restarting
PDM. 1F0002
Maximum number of
patient records + Oldest
record overwritten.
No data saved.
Save successful.
Possible cause
Clinician authentication failure
Clinician authentication failure
Clinician authentication failure
Clinician authentication failure
Clinician authentication failure
Clinician authentication failure
Clinician authentication failure
Patient authentication failure
The database was cleared due to
a schema update
The database was unreadable
during startup
Database corrupted when device
is in operation
Data was deleted as it contained
more than 400 records
A manual save is not allowed
A manual record was saved
Suggested action
Information status message; press OK button
to dismiss.
Information status message; press OK button
to dismiss.
Information status message; press OK button
to dismiss.
Information status message; press OK button
to dismiss.
Information status message; press OK button
to dismiss.
Information status message; press OK button
to dismiss.
Information status message; press OK button
to dismiss.
Information status message; press OK button
to dismiss.
Information status message; press OK button
to dismiss.
Press OK button to dismiss.
Press OK button to dismiss.
Information status message; press OK button
to dismiss.
Information status message; press OK button
to dismiss.
Information status message; press OK button
to dismiss.
Appendix 187
Alarm
priority
Information
Information
Information
Information
Information
Information
Information
Information
Information
Very low
Very low
Information
Information
Information