Advanced Settings; Access Advanced Settings - Welch Allyn Connex Spot Monitor Service Manual

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

Advanced settings

Advanced settings provide access to device settings and service menu information with
an access code. This section covers the following primary tabs:
General. Settings for system language as well as date and time, plus open source
software disclosure.
Parameters. Settings for physiological parameters, averaging, and intervals.
Network. Settings for the internal Ethernet and security, as well as WiFi radio and
Bluetooth subsystems.
Service. Settings and features that support installation and on-device
troubleshooting.
Note

Access Advanced settings

Note
1. From the Home tab, touch the Settings tab.
2. Touch the Advanced vertical tab.
3. Enter 6345 as the access code and touch OK.
Note
The General tab appears.
4. Perform service tasks by making selections or touching other tabs.
Note
5. When you are done, touch Exit.
The Home tab appears.
Changes made in Advanced settings take effect immediately, but they do
not change the configuration file and cannot be used to clone settings from
one device to another. Follow the instructions in "Device configuration and
setup" to create, edit, download, and install any settings you want available
on multiple devices.
You cannot access Advanced settings if sensors or physiological alarms are
active or if vital sign measurements are displayed.
This access code is configurable and could be different at
your location.
Service tasks and how to do them are detailed in this
section.
23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents