Full functional verification check and calibration tools
The list of tools below is what is required to perform a full device functionality check and
calibration. The tools are used in conjunction with the Welch Allyn service tool, Gold
license, to perform a device calibration.
Material no.
01802-110
407672
Material
No.
Description
Doc-10
Nellcor SpO2 extension cable
Varies
APM (configurations with Accessory
Power Management)
Varies
Power supply (configurations with no
APM)
106270
Service test box
106275
Micro USB cable
N/A
Pressure meter (must include at least
two decimal points and be accurate
to within ±0.5 mmHg)
407672
Test volume repair fixture (113670)
with test manifold, bulb, and valve
N/A
Blood pressure Y-tube
01802-110
Tester, calibration, 9600 Plus,
SureTemp
06138-000
Cal-key, assembly, M690/692
All items listed in this table are required to complete a full functional verification and calibration. If
Welch Allyn does not make an item available for purchase, you must provide it yourself.
You may substitute items marked as an optional Welch Allyn purchase with comparable OEM
products.
Description
Tester, calibration, 9600 Plus
BP test volume repair fixture 113670
Functional verification and calibration overview 123
Required Welch
Allyn purchase
(see Notes)
Qty
1
1
X
1
X
1
1
1
1
1
See instructions for creating the Y-tube in
the Basic functional verification section.
3
1
Optional Welch
Allyn purchase
X
X
X
X
X
X
Qty Component
3
Braun
ThermoScan
PRO 6000 &
SureTemp Plus
probes
1
NIBP