Repairs

Repairs should be performed at the factory.

Cleaning and Disinfection

Lenses: PanOptic is a precision optical instrument. Do not use any
solvent-based cleaners on the lenses. Use only a clean optical grade
cloth to clean the lenses.
Housing: Wipe the PanOptic housing with an appropriate health care
low- or intermediate-level cleaner/disinfecting wipe that incorporates
either a 1:10 sodium hypochlorite (bleach) solution or isopropyl alcohol
as the active disinfection ingredient. Follow wipe manufacturer's
instructions for appropriate use, contact times and applicable
warnings and precautions.
Do not excessively saturate the PanOptic.
Do not immerse PanOptic in any solution.
Do not sterilize the PanOptic.
After disinfection, inspect the PanOptic for visible signs of
deterioration. If evidence of damage or deterioration is present,
discontinue use and contact Welch Allyn or your area representative.
Patient Eyecup: Wipe the Patient Eyecup with an appropriate health
care low- or intermediate-level cleaner/disinfecting wipe that
incorporates isopropyl alcohol as the active disinfection ingredient or
soak in a glutaraldehyde-based solution. Follow product
manufacturer's instructions for appropriate use, contact times and
applicable warnings and precautions.
Patient Eyecup can be autoclaved.
Do not use acetone based products or other harsh chemicals.
No other preventative maintenance is necessary for this product.