120 Functional verification and calibration overview
Thermometry tests
Calibration key temperature test for SureTemp Plus
Temperature test
97.3 °F (36.3°C)
SureTemp Plus test
Temperature tested
96.8 °F (36.0 °C)
101.3 °F (38.5 °C)
105.8 °F (41.0 °C)
Braun ThermoScan PRO 6000 test
Temperature tested
96.8 °F (36.0 °C)
101.3 °F (38.5 °C)
105.8 °F (41.0 °C)
Specification ± 0.2°F (± 0.1°C)
97.1 to 97.5°F
36.2 to 36.4°C
Specification ± 0.2 °F (± 0.1 °C)
96.6 °F to 97.0 °F
35.9 °C to 36.1 °C
101.1 °F to 101.5 °F 38.4 °C to 38.6 °C
105.6 °F to 106.0 °F 40.9 °C to 41.1 °C
Specification ± 0.4 °F (± 0.2 °C)
96.4 °F to 97.2 °F
35.8 °C to 36.2 °C
100.9 °F to 101.7 °F 38.3 °C to 38.7 °C
105.4 °F to 106.2 °F 40.8 °C to 41.2 °C
®
Connex
Spot Monitor
Actual reading
Pass
°F or °C
Actual reading
Pass
°F or °C
°F or °C
°F or °C
Actual reading
Pass
°F or °C
°F or °C
°F or °C
Fail
Fail
Fail