Welch Allyn Connex Spot Monitor Service Manual page 203

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

Service manual
Message
CSM battery is not
installed. 1C100E
Device shutdown is not
available at this time
No valid files found
Firmware update
successful.
Audio alarms are off.
Advanced settings
unavailable.
USB accessory
disconnected.
Advanced settings
Save not successful.
Save successful.
Software upgrade is
downloading. Do not
shutdown.
Factory reset successful. The monitor was reset to factory
Factory reset failed.
Custom configuration
file not deleted.
Configuration upload
successful.
Unable to load
configuration; using
factory defaults.
Possible cause
There is no battery in the monitor Check the monitor for a battery and install
System shutdown failure
No valid files were found on the
USB flash drive
The software successfully updated Information status message; press OK
The monitor alarm audio is off
The advanced settings are
unavailable because the monitor
is not in an idle state
The USB device was disconnected
from the monitor
The Advanced settings code was
entered correctly
The device configuration or logs
were not saved to the USB device
The device configuration or logs
were saved to the USB device
The monitor is downloading a
software update
settings
The monitor factory reset has
failed.
The device configuration has
successfully uploaded
The device configuration was not
successfully uploaded
Suggested action
one if it is missing. If the problem is still
present, perform diagnostic checks on the
monitor. If the problem persists, replace the
main PCBA.
Monitor cannot perform an immediate
shutdown. Disconnect AC power and
remove battery.
Re-insert the USB flash drive with valid
files.
button to dismiss.
Information status message; press OK
button to dismiss.
Confirm that no sensors are attached to the
monitor, no alarms are active, and there is
no unsaved data in the Spot or Intervals
profile.
Information status message; press OK
button to dismiss
Information status message; Exit Advanced
settings to dismiss.
Information status message; press OK
button to dismiss
Information status message; press OK
button to dismiss
Information status message; press OK
button to dismiss
Information status message; press OK
button to dismiss
Information status message; Exit Advanced
settings to dismiss.
Information status message; press OK
button to dismiss
Information status message; Exit Advanced
settings to dismiss.
Appendix 197
Alarm
priority
Very low
Information
Information
Information
Information
Information
Information
Information
Information
Information
Information
Information
Information
Information
Information

Hide quick links:

Advertisement

Table of Contents
loading