Power Down The Monitor - Welch Allyn Connex Spot Monitor Service Manual

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

Service manual
Note

Power down the monitor

You can power down the device as follows: 1) by briefly pressing the power button on
the housing and then following the onscreen prompts, or 2) by using onscreen controls
alone.
Option 1. Press the power button. An onscreen dialog opens with the following options:
Power down. Device behavior is the same for all profiles.
Sleep. The Sleep button clears the display and puts the device into Display power-
saving mode.
Cancel. The Cancel button dismisses the dialog.
Touch Power down and complete the power-down procedure as described in Option 2,
steps 3 and 4.
Option 2. Use onscreen controls alone to power down the device:
1. Touch the Settings tab.
2. Touch the Device tab.
3. Touch Power down.
4. Select the desired option and then touch Power down.
The Power-down options dialog
The buttons in this dialog produce varied effects, as noted below:
Power down. The device clears all data onscreen and performs a complete software
shutdown.
Cancel. This button dismisses the dialog and returns you to the previous screen.
Note
After performing any of these procedures and before returning the device
to service, you must use the service tool, Gold edition, to complete the full
suite of functional tests to ensure that all systems are operating within the
design specifications. For more information about these tests and the
service tool, see "Functional verification and calibration." If you do not have
the service tool, you can download it from welchallyn.com.
If your monitor is configured with a Braun thermometer, avoid powering
down the monitor between patient measurements to help maintain an
adequate charge to the Braun dock.