Reassembly notes
SpO2 insert options are Masimo, Nellcor, Nonin, or none.
Kit items:
105926, Masimo insert only
105927, Nellcor insert only
105929, Nonin insert only
105928, No SpO2 insert only
Note
When replacing the SpO2 insert
1. First complete the following steps:
Attach the power harness to the SpO2 insert. (See "Attach the power harness to the
insert.")
Install any new SpO2 board, if applicable. (See the reassembly notes in "Remove the
SpO2 board." )
Attach the power harness to the insert
Align the power harness with the insert and secure it with 2 T-10 Torx screws (see
the Torque Value Table).
Reassembly is not the reverse of disassembly, but it is the same for each
insert.