Welch Allyn Connex Spot Monitor Service Manual page 118

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

112 Functional verification and calibration overview
Blood Pressure Bulb
Pressure Meter
Dual Lumen Pressure Tubing
Pressure Test Box
1. Navigate to the Advanced tab.
2. In the Advanced settings code box, enter 6345.
3. Navigate to Service > General.
The NIBP Pressure Verification feature is located in the bottom right of the screen.
4. Connect the dual lumen tubing to the device.
5. Connect one end of the tubing to the pressure meter and the other end to the test
box. Ensure each end is connected tightly and there are no leaks.
6. To initiate the verification process, touch Close Valve.
7. Use the blood pressure bulb to adjust the pressure to each threshold of
20/50/100/150/250 mmHg.
8. Compare the pressure displayed on the NIBP Pressure Verification screen and the
pressure meter.
SpO2 tests
Select the procedure here that applies to your configuration to test the device's SpO2
function. Connect the test equipment according to the test setup diagram for your
configuration presented here.
SpO2 test setup
For the SpO2 test, connect the test equipment shown below.
®
Connex
Spot Monitor