168 Appendix
Message
Replace the SpO2 cable.
040800
Replace the SpO2 cable.
040900
Replace the SpO2 cable.
040a00
Sensor not connected.
Clear to retry. 040b00
The sensor has expired.
Replace the SpO2
sensor. 040c00
Incompatible sensor.
Replace the SpO2
sensor. 040d00
Incompatible sensor.
Replace the SpO2
sensor. 040e00
Possible cause
The SpO2 cable is incompatible
with the monitor
The SpO2 cable is not recognized
by the monitor
The SpO2 cable is defective
The SpO2 sensor is not connected
to the monitor
The SpO2 sensor has expired
The SpO2 sensor is not recognized
by the monitor
The SpO2 sensor is not recognized Replace SpO2 sensor. If the problem
Suggested action
tester. If the message persists,
replace the module.
Replace the SpO2 cable. If the
problem is still present, verify
module functionality by replacing the
sensor with an applicable SpO2
tester. If the message persists,
replace the module.
Replace the SpO2 cable. If the
problem is still present, verify
module functionality by replacing the
sensor with an applicable SpO2
tester. If the message persists,
replace the module.
Replace the SpO2 cable. If the
problem is still present, verify
module functionality by replacing the
sensor with an applicable SpO2
tester. If the message persists,
replace the module.
Connect SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
®
Connex
Spot Monitor
Alarm priority
Very low
Very low
Very low
Very low
Very low
Very low
Very low