Welch Allyn Connex Spot Monitor Service Manual page 204

Hide thumbs Also See for Connex Spot Monitor:
Table of Contents

Advertisement

198 Appendix
Software update messages
Message
Software Update:
Manifest transfer timed
out. Verify connection
and retry.
Software Update:
Package file transfer
timed out. Verify
connection and retry
Software Update:
Invalid token file.
Software Update:
Unable to find manifest
file on server.
Software Update:
Unable to verify
manifest file signature.
Software Update:
Package file corrupted.
Regenerate package and
retry.
Software Update:
Unable to find package
file.
Software Update:
Installation failed.
Reboot and retry.
Software Update:
Upgrade unsuccessful.
Insufficient disk space.
Software Update:
Update unsuccessful.
Incompatible firmware.
Software Update: SWUP
internal error
Software Update:
Manager internal error
Radio software upgrade
failed. 350600
Possible cause
The manifest file transfer timed
out or connection was lost during
download
The package file transfer timed
out or connection was lost during
download
There was an invalid token file
The manifest file was not found
on the server
The manifest file signature
verification failed.
The package file is corrupt, does
not have expected SHA256 hash
The package file cannot be found Verify the package file is on the server.
At least one of the sub systems
failed to install
The partition is running out of
space
The current firmware version is
lower than the one required by the
token file
SWUP NIBP is not functional
The Software Update manager is
not functional
The radio software was not
upgraded.
Suggested action
Verify the connection and retry.
Verify the connection and retry.
Verify and update the token file.
Verify the manifest file is on the server.
Regenerate the software package and retry. Information
Regenerate the software package and retry. Information
Restart the monitor.
Free up adequate space needed to perform
the upgrade.
Try to update to an earlier software
package.
Information status message; click OK button
to dismiss.
Information status message; click OK button
to dismiss.
Check for software update and apply it. If
message is still present, replace the radio.
®
Connex
Spot Monitor
Alarm
priority
Information
Information
Information
Information
Information
Information
Information
Information
Information
Information
Very low

Hide quick links:

Advertisement

Table of Contents
loading