Article number
15.2 Operating states/error signals
The orthosis indicates operating states and error messages with symbols on the control panel and
through beeps and vibration signals.
15.2.1 Status display on the control panel
The current status of the component is displayed on the control panel for 5 seconds in the follow
ing cases:
•
The
button on the control panel was pressed briefly.
•
The component was switched on by pressing the
•
The battery charger was disconnected from the component.
•
Battery charger was connected to the component.
•
An error was identified during use.
INFORMATION
No status display due to battery deep discharge
If the status of the component is not displayed on the control panel, the battery may be deep dis
charged. In this case it must be charged for at least 15 minutes before the status can be queried
again.
Symbols
Beep signal
on the
control
panel
All sym
bols on
the con
trol panel
light up
altern
ately
94
Vibra
tion sig
nal
—
—
—
—
button.
Event
Test of the indicators
(LEDs) after discon
necting the battery
charger
The product is
switched on and oper
ational
Required action
Check whether all
symbols (LEDs) light
up alternately and in
the corresponding col
ours.
If a symbol (LED) does
not light up in a colour,
the product should be
inspected by an O&P
professional.
If no symbols (LEDs)
light up, the battery
may be deep dis
charged. Leave the
battery charger con
nected for at least
15 minutes and then
repeat this test by dis
connecting/connecting
the battery charger.