15.2 Operating states/error signals

The orthosis indicates operating states and error messages with symbols on the control panel and
through beeps and vibration signals.

15.2.1 Status display on the control panel

The current status of the component is displayed on the control panel for 5 seconds in the follow­
ing cases:
The
button on the control panel was pressed briefly.
The component was switched on by pressing the
The battery charger was disconnected from the component.
Battery charger was connected to the component.
An error was identified during use.
INFORMATION
No status display due to battery deep discharge
If the status of the component is not displayed on the control panel, the battery may be deep dis­
charged. In this case it must be charged for at least 15 minutes before the status can be queried
again.
Symbols
Beep signal
on the
control
panel
All sym­
bols on
the con­
trol panel
light up
altern­
ately
94
Vibra­
tion sig­
nal
button.
Event
Test of the indicators
(LEDs) after discon­
necting the battery
charger
The product is
switched on and oper­
ational
Required action
Check whether all
symbols (LEDs) light
up alternately and in
the corresponding col­
ours.
If a symbol (LED) does
not light up in a colour,
the product should be
inspected by an O&P
professional.
If no symbols (LEDs)
light up, the battery
may be deep dis­
charged. Leave the
battery charger con­
nected for at least
15 minutes and then
repeat this test by dis­
connecting/connecting
the battery charger.